Functional Check
Rev May 1994
WFM 601i Serial Component Monitor
1–23
.7
.6
.8
.5
.4
.3
.2
.1
–.1
–.2
–.3
Tek
COMPONENT ANALOG
2% & 4% K
PB
100%
50%
CAL SIG
ON OFF
Figure 1-17: WFM 601i calibrator display.
7. Gain Calibration
The WFM 601i has both factory preset and user adjustable horizontal and vertical
gain calibration. To access GAIN CAL (you must be in the
CONFIGURE/CALIBRATE submenu) push the second from the bottom bezel
button and observe that the cursor box now surrounds the ON readout.
The other two assignable bezel controls are now labeled V CAL and H CAL.
Rotate both controls enough to see that the horizontal and vertical gains have
changed noticeably and that two new labels have appeared next to the bezel
buttons. They are RESET V CAL and RESET H CAL. See Figure 1-18.
Push the RESET V CAL button to return to the factory calibration setting. Note
that the vertical amplitude returns to 700 mV and that the RESET V CAL label
disappears.
Push the RESET H CAL and note that the horizontal gain returns to one cycle
per division and the RESET H CAL label disappears. Exit the configure menu
by pressing the CONFIG button. The calibrator signal is automatically turned
off.
Summary of Contents for WFM 601i
Page 4: ......
Page 12: ...Contents vi WFM 601i Serial Component Monitor...
Page 16: ...Preface x WFM 601i Serial Component Monitor...
Page 17: ...Getting Started...
Page 18: ......
Page 51: ...Operating Basics...
Page 52: ......
Page 58: ...At A Glance 2 6 WFM 601i Serial Component Monitor...
Page 76: ...Functional Overview 2 24 WFM 601i Serial Component Monitor...
Page 77: ...Reference...
Page 78: ......
Page 104: ...Measurement Theory 3 26 WFM 601i Serial Component Monitor...
Page 105: ...Appendices...
Page 106: ......
Page 118: ...Appendix B Multipin Connectors B 4 WFM601i Serial Component Monitor...
Page 123: ...Glossary...
Page 124: ......
Page 129: ...Index...
Page 130: ......
Page 134: ......
Page 135: ......