Appendix C: User Service
WFM601i Serial Component Monitor
C–4
Replacing the CRT Filter
A smoke-gray filter is installed over the face of the CRT. If the filter becomes
damaged, it can be replaced in the following manner:
1. Remove the five knobs below the CRT.
2. Insert a small, straight-blade screwdriver into the recessed area on either the
right or left side of the panel. Pry gently until the panel snaps out of the
front-panel frame. See Figure C-1 for panel and recess location.
3. Remove the two Torx
screws that fasten the bezel to the instrument.
4. Grasping the bottom of the bezel, pull out and upward to remove the bezel
from the instrument. (There are two hinges at the top of the bezel that hold it
in place; once the bezel is at an approximate 45
°
angle with the front panel,
they will disengage.)
5. Remove the damaged filter and snap the new filter into place on the back
side of the bezel. Position the ridged side of the filter towards the CRT to
eliminate unwanted visual effects on the CRT face.
6. Replace the bezel and bezel screws. Tighten screws (8 in/lbs).
7. Replace the panel below the CRT, and press on both the right and left sides
of the panel until it snaps into place.
8. Replace the five knobs below the CRT and tighten the set screws.
Summary of Contents for WFM 601i
Page 4: ......
Page 12: ...Contents vi WFM 601i Serial Component Monitor...
Page 16: ...Preface x WFM 601i Serial Component Monitor...
Page 17: ...Getting Started...
Page 18: ......
Page 51: ...Operating Basics...
Page 52: ......
Page 58: ...At A Glance 2 6 WFM 601i Serial Component Monitor...
Page 76: ...Functional Overview 2 24 WFM 601i Serial Component Monitor...
Page 77: ...Reference...
Page 78: ......
Page 104: ...Measurement Theory 3 26 WFM 601i Serial Component Monitor...
Page 105: ...Appendices...
Page 106: ......
Page 118: ...Appendix B Multipin Connectors B 4 WFM601i Serial Component Monitor...
Page 123: ...Glossary...
Page 124: ......
Page 129: ...Index...
Page 130: ......
Page 134: ......
Page 135: ......