Functional Check
1–18
WFM 601i Serial Component Monitor
Initial Equipment Connections
H
Connect the WFM 601i to an appropriate AC power source.
H
Connect the component serial digital signal to SER A and terminate the other
side of the loop-through with a 75
W
terminator.
This procedure does not check at the Monitor Out, but if there is a GBR or
Y–P
B
–P
R
analog component color monitor available, it can be hooked up to
check these outputs.
Procedure
1. Initialize the Front-Panel Controls
The TektronixWFM 601i Serial Component Monitor is shipped with an internal
preset which sets the front-panel controls to a factory-defined setting. To reset
the front-panel controls to the factory presets, enter the MENU by pressing the
PRESET button. Select the FACTORY setting by turning the front-panel bezel
knob under the list of presets until FACTORY is highlighted. Press the bezel
button adjacent to RECALL. The PRESET MENU is automatically exited, and
the front-panel control settings are set to the factory preset. This preset
automatically selects the line rate (625 or 525). Note that upper left hand corner
of the CRT displays the scan rate with a readout preceded by an
A
(scan rate
automatically selected.) Check that the instrument is now in the Waveform
display mode with the luminance portion of the Color Bar signal displayed at a
line sweep of 10
m
s/Div.
2. Display Control
Adjust the VERT and HORIZ POS knobs to center the signal display. The CRT
menu allows the operator to adjust the instrument CRT display. Enter the menu
by pressing the
CRT
button. The knobs under the CRT now control the focus,
scale illumination, and display intensity of the signal. Adjust the controls to the
desired viewing levels. See Figure 1-12.
Summary of Contents for WFM 601i
Page 4: ......
Page 12: ...Contents vi WFM 601i Serial Component Monitor...
Page 16: ...Preface x WFM 601i Serial Component Monitor...
Page 17: ...Getting Started...
Page 18: ......
Page 51: ...Operating Basics...
Page 52: ......
Page 58: ...At A Glance 2 6 WFM 601i Serial Component Monitor...
Page 76: ...Functional Overview 2 24 WFM 601i Serial Component Monitor...
Page 77: ...Reference...
Page 78: ......
Page 104: ...Measurement Theory 3 26 WFM 601i Serial Component Monitor...
Page 105: ...Appendices...
Page 106: ......
Page 118: ...Appendix B Multipin Connectors B 4 WFM601i Serial Component Monitor...
Page 123: ...Glossary...
Page 124: ......
Page 129: ...Index...
Page 130: ......
Page 134: ......
Page 135: ......