Reference
Menu Off
Push
Menu Off
to clear displayed menus from the screen.
Print-Ready Screenshots
You can set up the oscilloscope to save print-ready images through the
Utility
►
- more - page 1 of 2
(page 1 of 3 on EDU models)
►
Options
►
Printer
Setup
menu.
Option
Setting
Comments
Ink Saver
On, Off
Prints the screen image on a
white background when you
select On
Layout
1
Portrait, Landscape
Printer output orientation
File Format
Bmp, Jpg
fi
le format
1
The printer may override your selection for best
fi
t.
Reference Menu
The Reference menu can turn on or turn off reference memory waveforms
from the display. The waveforms are stored in the non-volatile memory of the
oscilloscope, and have the following designations: RefA and RefB.
To display (recall) or hide a reference waveform, follow these steps:
1.
Push the
Ref
front panel button.
2.
Push the side-menu button corresponding to the reference waveform you
wish to display or hide.
Reference waveforms have the following characteristics:
Reference waveforms are displayed in white
Two reference waveforms can be displayed at the same time
Vertical and horizontal scale readouts display at the bottom of the screen
Reference waveforms cannot be zoomed or panned
You can display one or two reference waveforms at the same time as "live"
channel waveforms. If you display two reference waveforms, you must hide one
waveform before you can display a different one.
Refer to
Save Waveform
for information on how to save reference waveforms.