Getting Started
Manual Probe Compensation
You can manually perform this adjustment to match your probe to the input
channel.
PROBE COMP
Autoset
button
1.
Push the
1
►
Probe
►
Voltage
►
Attenuation
option and select
10X
. Connect the appropriate
TPP0051, TPP0101, or TPP0201 probe to
channel 1 on the oscilloscope. If you use the
probe hook-tip, ensure a proper connection by
fi
rmly inserting the tip onto the probe.
2.
Attach the probe tip to the PROBE COMP
~5V@1kHz terminal and the reference lead to
the PROBE COMP chassis terminal. Display the
channel, and then push the
Autoset
button.
Overcompensated
Undercompensated
Compensated correctly
3.
Check the shape of the displayed waveform.
4.
If necessary, adjust your probe.
Repeat as necessary.
Probe Attenuation Setting
Probes are available with various attenuation factors which affect the vertical
scale of the signal.
Select the factor that matches the attenuation of your probe. For example, to
match a probe set to 10X connected to CH 1, push the
1
►
Probe
►
Voltage
►
Attenuation
option, and select
10X
.
NOTE.
The default setting for the Attenuation option is 10X.