Reference
The following conditions cause autorange to adjust settings:
Too many or too few waveform periods for a clear display of the trigger
source (except when in Vertical Only)
Waveform amplitude too large or too small (except when in Horizontal Only)
Ideal trigger level changes
When you push the
Autoset
button for more than 1.5 seconds, the oscilloscope
enters the autorange mode and adjusts controls to produce a usable display of
the input signal.
Function
Setting
Acquire mode
Sample
Display format
YT
Display persist
Off
Horizontal position
Adjusted
Horizontal view
Main
Run/Stop
RUN
Horizontal scale (seconds/division)
Adjusted
Trigger coupling
DC
Trigger holdoff
Minimum
Trigger level
Adjusted
Trigger mode
Edge
Vertical bandwidth
Full
Vertical BW limit
Off
Vertical coupling
DC
Vertical invert
Off
Vertical scale (volts/division)
Adjusted
The following changes to the setup of the oscilloscope deactivate autorange:
Vertical scale deactivates vertical autoranging
Horizontal scale deactivates horizontal autoranging
Display or remove a channel waveform
Trigger settings
Single sequence acquisition mode
Recall a setup
XY Display format
Persistence