Functional Check
Perform this functional check to verify that your oscilloscope is operating
correctly.
ON/OFF button
1.
Power on the oscilloscope.
Push the
Default Setup
button.
The default Probe option attenuation setting is
10X.
Default Setup button
PROBE COMP
2.
Connect the TPP0051, TPP0101, or TP0201
probe to channel 1 on the oscilloscope. To do this,
align the slot in the probe connector with the key
on the channel 1 BNC, push to connect, and twist
to the right to lock the probe in place.
Connect the probe tip and reference lead to the
PROBE COMP terminals.
3.
Push the
Autoset
button. Within a few seconds,
you should see a square wave in the display of
about 5 V peak-to-peak at 1 kHz.
Push the channel
1
menu button on the front panel
twice to remove channel 1, push the channel
2
menu button to display channel 2, and repeat
steps 2 and 3.
Calibration: PASSED
4.
Check that the instruction passed the calibration
tests. Push
Utility
►
- more - page 1 of 2
(push
more
again on EDU models)
►
System Status
►
Misc.
. Look for
Calibration PASSED
Security cable channel
Security lock hole
Power cord
Ventilation
NOTE.
The oscilloscope cools by convection. Keep two inches clear on the sides
and top of the product to allow adequate air
fl
ow.