Application Examples
8.
Turn the
Multipurpose
knob to highlight
Ch1
.
9.
Push the
Multipurpose
knob to select
Ch1
.
10.
Push the
Cursor 1
option button.
11.
Push the
Cursor 1
option button.
12.
Turn the
Multipurpose
knob to place a cursor on the active edge of the
chip-select signal.
13.
Push the
Cursor 2
option button.
14.
Turn the
Multipurpose
knob to place the second cursor on the data output
transition.
The
Δ
t
readout in the Cursor Menu is the propagation delay between the
waveforms. The readout is valid because the two waveforms have the same
horizontal scale (seconds/division) setting.
Triggering on a Speci
fi
c Pulse Width
To set up a test for pulse width aberrations, follow these steps:
1.
Push the
Autoset
button to trigger a stable display.
2.
Push the single cycle
option button in the Autoset Menu to view a
single cycle of the signal, and to quickly take a Pulse Width measurement.
3.
Push the
Trigger Menu
button to see the Trigger Menu.
4.
Push
Type
.
5.
Turn the
Multipurpose
knob to highlight
Pulse
from the pop-out menu. Push
the knob to select the choice
6.
Push
Source
.
7.
Turn the
Multipurpose
knob to highlight
Ch1
from the pop-out menu. Push
the knob to select the choice
w w w . e k t