Appendix B: Automatic Measurements Reference
B- 8
CSA8000B & TDS8000B User Manual
Pulse Measurements - Timing
Table B--2 describes each pulse measurement in the timing category. See Table
on B--1 on page B--2 for amplitude category measurements; see Table B--3 on
page B--14 for area category measurements.
Table B- 2: Pulse Measurements - Timing
Name
Definition
Burst width
The time between the first and last crossings, either positive or negative, of the waveform at the
mid-reference level in the measurement region. See
Measurement Reference Parameters and
Methods
on page B--56 or in the online help. Also see
Reference Levels Method
on page 3--79.
The mid-reference level is adjustable and defaults to 50% of the pulse amplitude.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See
To Localize a Measurement
on page 3--83.
Cross+
The time of the first positive crossing of the data sampled at the mid-reference level in the
measurement region.
Cross+ = Tcross
Where Tcross is the horizontal coordinate of the first positive crossing. See
Pulse Sources
on page B--57.
The mid-reference level is adjustable and defaults to 50% of the pulse amplitude.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). By default, searching is forward from the Start Gate for the first
rising edge, but can be reversed, so that searching is backward from the Stop Gate.
Cross-
The time of the first negative crossing of the data sampled at the mid-reference level in the
measurement region.
Cross-- = Tcross
Where Tcross is the horizontal coordinate of the first negative crossing.
The mid-reference level is adjustable and defaults to 50% of the pulse amplitude.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). By default, searching is forward from the Start Gate for the first
falling edge, but can be reversed, so that searching is backward from the Stop Gate.
Summary of Contents for CSA8000B Series
Page 4: ......
Page 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Page 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Page 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Page 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Page 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Page 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Page 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Page 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Page 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Page 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Page 353: ......
Page 354: ......