Appendix B: Automatic Measurements Reference
B- 22
CSA8000B & TDS8000B User Manual
Table B- 4: RZ Measurements - Amplitude (cont.)
Name
Definition
RZ Min
The minimum vertical value of the waveform that is sampled within the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
When this measurement is turned on, it will automatically set the measurement system to use a
waveform database if available.
For best results with this measurement:
H
Perform a Dark Level compensation before taking this measurement if the source of the
measured waveform is an optical channel. See
To Perform Dark-Level and User
Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
To Optimize the
Vertical Resolution
on page B--69.
RZ Peak-to-Peak
The difference between the Max and Min vertical values of the waveform that is sampled within
the measurement region.
RZ Peak-to-Peak = Max – Min
Where Max and Min are the maximum and minimum measurements.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See
To Localize a Measurement
on page 3--83.
When this measurement is turned on, it will automatically set the measurement system to use a
waveform database if available. See
Use a Waveform Database
on page B--70.
For best results with this measurement:
H
Perform a Dark Level compensation before taking this measurement if the source of the
measured waveform is an optical channel. See
To Perform Dark-Level and User
Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
How to Optimize the
Vertical Resolution
on page B--70.
Summary of Contents for CSA8000B Series
Page 4: ......
Page 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Page 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Page 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Page 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Page 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Page 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Page 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Page 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Page 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Page 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Page 353: ......
Page 354: ......