Appendix B: Automatic Measurements Reference
B- 40
CSA8000B & TDS8000B User Manual
Table B- 7: NRZ Measurements - Amplitude (cont.)
Name
Definition
NRZ Extinction Ratio
The ratio of the average power levels of the logic 1 level (High) to the logic 0 level (Low) of an
optical NRZ signal. All level determinations are made within the NRZ Eye Aperture.
Where High and Low are the logical 1 and 0 levels.
The Eye Aperture is adjustable and defaults to 20% of the NRZ bit time. See
RZ Eye Aperture
Parameters
on B--62.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See
To Localize a Measurement
on page 3--83.
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See
Use a Waveform Database
on page B--70.
For best measurement results:
H
Perform a Dark Level compensation before taking this measurement. See
To Perform
Dark-Level and User Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
How to Optimize the
Vertical Resolution
on page B--70.
NRZ ExtRatio
=
100
×
Ꮛ
Low
High
Ꮠ
NRZ Extinction Ratio (%)
The ratio of the average power levels of the logic 0 level (Low) to the logic 1 level (High) of an
optical NRZ signal, expressed as a percentage. All level determinations are made within the
NRZ Eye Aperture.
Where High and Low are the logical 1 and 0 levels.
The Eye Aperture is adjustable and defaults to 20% of the NRZ bit time.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
For best measurement results:
H
Perform a Dark Level compensation before taking this measurement. See
To Perform
Dark-Level and Gain User Wavelength Gain Compensations
on page 3--98.
H
Optimize the vertical resolution before taking this measurement. See
To Optimize the
Vertical Resolution
on page B--69.
NRZ ExtRatio
[%]
=
100
×
Ꮛ
Low
High
Ꮠ
Summary of Contents for CSA8000B Series
Page 4: ......
Page 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Page 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Page 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Page 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Page 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Page 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Page 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Page 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Page 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Page 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Page 353: ......
Page 354: ......