Appendix B: Automatic Measurements Reference
CSA8000B & TDS8000B User Manual
B- 7
Table B- 1: Pulse Measurements — Amplitude (cont.)
Name
Definition
Peak-to-Peak Noise
The maximum range of the waveform amplitude variance sampled within a fixed width vertical
slice located at the center of the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See
To Localize a Measurement
on page 3--83.
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See
Use a Waveform Database
on page B--69.
For best results with this measurement. optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
Pk-Pk
The absolute difference between the maximum and minimum amplitude values of the waveform
in the measurement region. See example on page B--63.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
RMS
The true root-mean-square value of the waveform over the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
RMS Noise
One standard deviation of the waveform amplitude variance, sampled within a fixed-width
vertical slice located at the center of the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
For best results with this measurement, optimize the vertical resolution before taking this
measurement. See
To Optimize the Vertical Resolution
on page B--69.
Signal-to-Noise Ratio
The ratio of the signal amplitude to the noise level. The noise level is defined as one standard
deviation of the waveform amplitude variance within a fixed width vertical slice located at the
center of the measurement region.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
Summary of Contents for CSA8000B Series
Page 4: ......
Page 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Page 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Page 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Page 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Page 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Page 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Page 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Page 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Page 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Page 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Page 353: ......
Page 354: ......