Acquiring Waveforms
3- 6
CSA8000B & TDS8000B User Manual
CAUTION.
Install sampling modules before applying power and before connect-
ing them to the signals you want to test. See your sampling-module user manual
for instructions.
CAUTION.
Sampling modules are inherently vulnerable to static damage. Always
observe static-safe procedures and cautions as outlined in your sampling-module
user manual
.
Coupling Concerns.
Electrical sampling modules provide only straight-DC
coupling to their sampling circuits, with no protection. All modules specify a
maximum vertical nondestructive range that limits signals to small levels,
typically about 2 to 3 volts (DC + AC
pk-pk
). (See
Specifications
in the user
manual for your sampling module for exact limits.) Do not exceed the limit, even
momentarily, as the input channel may be damaged.
All modules also specify a dynamic range that, if exceeded, could cause
acquisition and measurement errors due to nonlinearity. Do not exceed this limit.
(See
Specifications
in the user manual for your sampling module for exact
limits.)
NOTE
. Optical sampling modules may have dynamic range exceeded without
obvious visual indications onscreen because the photo detector and/or filters
used may not necessarily be able to pass through overloaded signals to the
sampler.
Use external attenuators if necessary to prevent exceeding the limits just
described. Note that there are no hardware bandwidth filters in most sampling
modules or in the instrument. (Some optical sampling modules have bandwidth
filters settable from the Vertical Setup menu of the instrument. See the user
manual for your optical sampling module for more information.)
Scaling, Offset, and Positioning Considerations.
These key controls determine the
portion of the input signal presented to the acquisition system:
H
Set the vertical offset to display the features of interest on your waveform
and avoid clipping. (See
Note
that follows.) Adjust the display control
Vertical Scale to control the portion of the vertical window displayed on
screen; adjust the display control Vertical Position to position the waveform
on screen. Note that vertical offset affects the vertical acquisition window,
but vertical scale and position do not. These last two controls are display
controls only.
Vertical Acquisition Window Considerations
on page 3--14
describes the vertical acquisition window.
Summary of Contents for CSA8000B Series
Page 4: ......
Page 18: ...Preface xiv CSA8000B TDS8000B User Manual ...
Page 26: ...Check the Package Contents 1 8 CSA8000B TDS8000B User Manual ...
Page 62: ...Accessories and Options 1 44 CSA8000B TDS8000B User Manual ...
Page 76: ...Overview 3 2 CSA8000B TDS8000B User Manual ...
Page 112: ...Acquiring Waveforms 3 38 CSA8000B TDS8000B User Manual ...
Page 126: ...Triggering 3 52 CSA8000B TDS8000B User Manual ...
Page 146: ...Displaying Waveforms 3 72 CSA8000B TDS8000B User Manual ...
Page 174: ...Measuring Waveforms 3 100 CSA8000B TDS8000B User Manual ...
Page 186: ...Creating Math Waveforms 3 112 CSA8000B TDS8000B User Manual ...
Page 214: ...Data Input and Output 3 140 CSA8000B TDS8000B User Manual ...
Page 353: ......
Page 354: ......