1730–Series Operating Instructions
2–23
HAD that exceeds 18
m
s, simply measure the bar top in increments by position-
ing the bar to the left or right from the leading or trailing edge. Note that when
the leading or trailing edge is on the appropriate arrow, the first or last 1
m
s is
automatically excluded from the measurement.)
Pulse–to–bar K–Factor measurements are made using the solid and short dashed
lines to the left of the line time distortion structure. These lines are scaled
according to the following formulas:
1 and 1
(1–4K) (1+4K)
Where:
K=0.02 for 2% K–Factor (dashed lines)
K=0.04 for 4% K–Factor (dashed lines)
Calibrated 5X Gain increases resolution to 0.4% and 0.8%.
This scaling is described in detail in CCIR Standard Volume 5, 1966.
Make sure that the center of the bar is at 100 IRE when blanking level is at
0 IRE (use VAR to adjust gain, if necessary). If necessary, use the HORIZON-
TAL Position control to place the 2T pulse over the measurement area and
measure its amplitude.
The Horizontal reference line is the baseline at 0 IRE (NTSC and PAL–M) or
0.3 V (PAL). This timing line is 12 divisions long on NTSC graticules (12.4
divisions for PAL), and takes on different timing intervals depending on the
sweep rate selected. In 2 line sweep each major division is 10
m
s, and when
magnified (X10), each major division equals 1
m
s. In 1 line sweep each major
division is equal to 5
m
s, and when magnified (X25) each major division equals
0.2
m
s. In 2 field sweep the timing scale is of no real value, since this is a
monitoring mode; however when 2 field sweep is magnified (X25) the entire
vertical (field) interval can be displayed.
The PAL graticule scales are from 0 to 1.2 V on the left side. See Figure 2-17.
The right side has markings at sync tip (–0.3 V), baseline (0 V), and peak white
(+0.7 V). There are 2% and 4% markings at the horizontal center of the graticule
on the 0 V line (sync tip level) to assist in measuring sync amplitude. The
dashed horizontal line at the top of the graticule is equal to 1.234 V to indicate
peak amplitude of 100% color bars.
Horizontal Scales for
NTSC and PAL Graticules
PAL Graticule Vertical
Scales
Summary of Contents for 1730 Series
Page 4: ......
Page 12: ...Contents viii...
Page 17: ...Introduction and Specifications...
Page 18: ......
Page 32: ...1730 Series Introduction 1 14...
Page 33: ...Operating Instructions...
Page 34: ......
Page 62: ...1730 Series Operating Instructions 2 28...
Page 64: ...Service Safety Summary S 2 1730 Series B070000 Above...
Page 66: ......
Page 67: ...Installation...
Page 68: ......
Page 82: ...1730 Series Installation 3 14...
Page 83: ...Theory of Operation...
Page 84: ......
Page 115: ...Checks and Adjustments...
Page 116: ......
Page 159: ...Maintenance...
Page 160: ......
Page 180: ...1730 Series Maintenance 6 20 3 Remove the board by slipping it through the front panel opening...
Page 184: ...1730 Series Maintenance 6 24 Figure 6 8 Repackaging a 1730 Series instrument...
Page 185: ...Options...
Page 186: ......
Page 189: ...Replaceable Electrical Parts...
Page 190: ......
Page 222: ...1730 Series Replaceable Electrical Parts 8 32...
Page 223: ...Diagrams Circuit Board Illustrations...
Page 224: ......
Page 246: ......
Page 247: ...Replaceable Mechanical Parts...
Page 248: ......
Page 255: ...1730 Series Waveform Monitor FIG 1 EXPLODED VIEW A1 A3 A2 A3A1 A10 A11...
Page 256: ...1730 Series B070000 Up...
Page 257: ......
Page 258: ......