1730–Series Checks and Adjustments
5–26
v. Check that the 1730–Series inputs are not terminated and that there is no
signal applied. Connect the Device Under Test connector on the RF
Bridge to the PIX MON OUT connector.
w. CHECK – that the return loss of the PIX MON OUT is better than
30 dB (15.8 mV), from 50 kHz to 6 MHz. Make this measurement with
instrument power on and no signal output.
End of Performance Check Procedure.
Summary of Contents for 1730 Series
Page 4: ......
Page 12: ...Contents viii...
Page 17: ...Introduction and Specifications...
Page 18: ......
Page 32: ...1730 Series Introduction 1 14...
Page 33: ...Operating Instructions...
Page 34: ......
Page 62: ...1730 Series Operating Instructions 2 28...
Page 64: ...Service Safety Summary S 2 1730 Series B070000 Above...
Page 66: ......
Page 67: ...Installation...
Page 68: ......
Page 82: ...1730 Series Installation 3 14...
Page 83: ...Theory of Operation...
Page 84: ......
Page 115: ...Checks and Adjustments...
Page 116: ......
Page 159: ...Maintenance...
Page 160: ......
Page 180: ...1730 Series Maintenance 6 20 3 Remove the board by slipping it through the front panel opening...
Page 184: ...1730 Series Maintenance 6 24 Figure 6 8 Repackaging a 1730 Series instrument...
Page 185: ...Options...
Page 186: ......
Page 189: ...Replaceable Electrical Parts...
Page 190: ......
Page 222: ...1730 Series Replaceable Electrical Parts 8 32...
Page 223: ...Diagrams Circuit Board Illustrations...
Page 224: ......
Page 246: ......
Page 247: ...Replaceable Mechanical Parts...
Page 248: ......
Page 255: ...1730 Series Waveform Monitor FIG 1 EXPLODED VIEW A1 A3 A2 A3A1 A10 A11...
Page 256: ...1730 Series B070000 Up...
Page 257: ......
Page 258: ......