1730–Series Maintenance
6–7
The following procedure is designed to assist in isolating problems, which in
turn expedites repairs and minimizes down time.
1. Ensure that the malfunction exists in the instrument. This is done by
making sure that the instrument is operating as intended by Tektronix (see
Operating Instructions in Section 2), and by checking that a malfunction has
not occurred up stream from the waveform monitor.
2. Determine and evaluate all trouble symptoms. This is accomplished by
isolating the problem to a general area such as an assembly. The block
diagram is a valuable aid in signal tracing and circuit isolation.
CAUTION. Probes and Meter Leads:
3. Determine the nature of the problem. Attempt to make the determina-
tion of whether the instrument is out of calibration or if there has been a
component failure. Once the type of failure has been determined, proceed on
to identify the functional area most likely at fault.
4. Visually inspect the suspect assembly for obvious defects. Most
commonly these will be broken or loose components, improperly seated
components, overheated or burned components, chafed insulation, etc.
Repair or replace all obvious defects.
5. Use successive electrical checks to locate the source of the problem. The
primary tool for problem isolation is the oscilloscope. Use the Performance
Check Procedure (located in Section 5) to determine if a circuit is operating
within specifications. At times it may be necessary to change a calibration
adjustment to determine if a circuit is operational, but since this can destroy
instrument calibration, care should be exercised. Before changing an
adjustment, note its position so that it can be returned to its original setting.
6. Determine the extent of the repair. If the necessary repair is complex, it
may be advisable to contact your local Tektronix field office or representa-
tive before continuing. If the repair is minor, such as replacing a component,
see the parts list for replacement information. Removal and replacement
procedures for the assemblies can be found under Corrective Maintenance.
General Troubleshooting
Techniques
Summary of Contents for 1730 Series
Page 4: ......
Page 12: ...Contents viii...
Page 17: ...Introduction and Specifications...
Page 18: ......
Page 32: ...1730 Series Introduction 1 14...
Page 33: ...Operating Instructions...
Page 34: ......
Page 62: ...1730 Series Operating Instructions 2 28...
Page 64: ...Service Safety Summary S 2 1730 Series B070000 Above...
Page 66: ......
Page 67: ...Installation...
Page 68: ......
Page 82: ...1730 Series Installation 3 14...
Page 83: ...Theory of Operation...
Page 84: ......
Page 115: ...Checks and Adjustments...
Page 116: ......
Page 159: ...Maintenance...
Page 160: ......
Page 180: ...1730 Series Maintenance 6 20 3 Remove the board by slipping it through the front panel opening...
Page 184: ...1730 Series Maintenance 6 24 Figure 6 8 Repackaging a 1730 Series instrument...
Page 185: ...Options...
Page 186: ......
Page 189: ...Replaceable Electrical Parts...
Page 190: ......
Page 222: ...1730 Series Replaceable Electrical Parts 8 32...
Page 223: ...Diagrams Circuit Board Illustrations...
Page 224: ......
Page 246: ......
Page 247: ...Replaceable Mechanical Parts...
Page 248: ......
Page 255: ...1730 Series Waveform Monitor FIG 1 EXPLODED VIEW A1 A3 A2 A3A1 A10 A11...
Page 256: ...1730 Series B070000 Up...
Page 257: ......
Page 258: ......