1730–Series Maintenance
6–18
3. To install a bulb, hold it with the tweezers as described in step 2,
position it in front of the socket, and push the bulb with your finger until it
snaps into place.
4. Replace the bezel.
1. Remove the bezel.
WARNING. The CRT may retain a dangerous charge. Ground the conductor of
the anode to discharge the CRT. Do not allow the conductor to touch your body
or any circuitry.
2. Slide a screwdriver with an insulated handle under the anode cap on the side
of the CRT, and ground the anode to the chassis, to discharge the CRT. DO
NOT touch the metal shaft of the screwdriver while doing this. Disconnect
the anode cap by prying it gently away from the CRT.
3. Disconnect J225 (trace rotation) on the Main board and push the connector
through the hole in the board.
4. The CRT can now be pulled straight out (some pressure may be needed).
The CRT shield, along with the grommet around its front and the rubber
manchet around its back, should come out with the CRT.
1. If the CRT is to be replaced, remove the metal shield from the neck of the
old CRT and place it around the neck of the new CRT, with the WARNING
sticker towards the top of the instrument. This should place the opening in
the grommet on the front edge of the shield towards the bottom of the
instrument. Ensure that the rubber manchet is on the back edge of the shield.
2. Slip the CRT part way back into position and feed the trace rotation wires
(and plug) back through the hole in the Main board.
WARNING. The CRT may retain a dangerous charge. Ground the the anode
connector to discharge the CRT. Do not allow the conductor to touch your body
or any circuitry.
3. Use a screwdriver to ground the anode connector on the CRT to the chassis.
4. Slide the CRT into the instrument, guiding the rubber manchet on the end of
the shield into the rear CRT support.
5. Align the socket on the A10 CRT Socket board with the pins and key on the
CRT. Gently push the CRT and the socket board together until the CRT pins
are fully seated in the socket.
CRT Removal
CRT Replacment
Summary of Contents for 1730 Series
Page 4: ......
Page 12: ...Contents viii...
Page 17: ...Introduction and Specifications...
Page 18: ......
Page 32: ...1730 Series Introduction 1 14...
Page 33: ...Operating Instructions...
Page 34: ......
Page 62: ...1730 Series Operating Instructions 2 28...
Page 64: ...Service Safety Summary S 2 1730 Series B070000 Above...
Page 66: ......
Page 67: ...Installation...
Page 68: ......
Page 82: ...1730 Series Installation 3 14...
Page 83: ...Theory of Operation...
Page 84: ......
Page 115: ...Checks and Adjustments...
Page 116: ......
Page 159: ...Maintenance...
Page 160: ......
Page 180: ...1730 Series Maintenance 6 20 3 Remove the board by slipping it through the front panel opening...
Page 184: ...1730 Series Maintenance 6 24 Figure 6 8 Repackaging a 1730 Series instrument...
Page 185: ...Options...
Page 186: ......
Page 189: ...Replaceable Electrical Parts...
Page 190: ......
Page 222: ...1730 Series Replaceable Electrical Parts 8 32...
Page 223: ...Diagrams Circuit Board Illustrations...
Page 224: ......
Page 246: ......
Page 247: ...Replaceable Mechanical Parts...
Page 248: ......
Page 255: ...1730 Series Waveform Monitor FIG 1 EXPLODED VIEW A1 A3 A2 A3A1 A10 A11...
Page 256: ...1730 Series B070000 Up...
Page 257: ......
Page 258: ......