1730–Series Maintenance
6–3
Table 6–1: (Cont.)Static Susceptibility
Relative Susceptibility Levels
Voltage
LOW POWER SCHOTT-
KY TTL
TTL
Observe the following precautions to avoid damage:
1. Minimize handling of static–sensitive components.
2. Transport and store static–sensitive components or assemblies in their
original containers, on a metal rail, or on conductive foam. Label any
package that contains static–sensitive components or assemblies.
3. Discharge the static voltage from your body, by wearing a wrist
grounding strap, while handling these components. Servicing static–sensi-
tive assemblies or components should be done only at a static–free work
station by qualified personnel.
4. Nothing capable of generating or holding a static charge should be
allowed on the work station surface.
5. Keep the component leads shorted together whenever possible.
6. Pick up the components by the body, never by the leads.
7. Do not slide the components over any surface.
8. Avoid handling components in areas that have a floor or work surface
covering capable of generating a static charge.
9. Use a soldering iron that is connected to earth ground.
10. Use only special antistatic, suction, or wick–type desoldering tools.
Instrument performance should be checked after each 2000 hours of operation, or
every 12 months, if used intermittently. This will help to ensure maximum
performance and assist in locating defects that may not be apparent during
regular operation. The Performance Check Procedure and the Adjustment
Procedure are in Section 5.
Troubleshooting
The material contained here is general and is not intended to cover specific cases.
Performance Checks and
Readjustments
Summary of Contents for 1730 Series
Page 4: ......
Page 12: ...Contents viii...
Page 17: ...Introduction and Specifications...
Page 18: ......
Page 32: ...1730 Series Introduction 1 14...
Page 33: ...Operating Instructions...
Page 34: ......
Page 62: ...1730 Series Operating Instructions 2 28...
Page 64: ...Service Safety Summary S 2 1730 Series B070000 Above...
Page 66: ......
Page 67: ...Installation...
Page 68: ......
Page 82: ...1730 Series Installation 3 14...
Page 83: ...Theory of Operation...
Page 84: ......
Page 115: ...Checks and Adjustments...
Page 116: ......
Page 159: ...Maintenance...
Page 160: ......
Page 180: ...1730 Series Maintenance 6 20 3 Remove the board by slipping it through the front panel opening...
Page 184: ...1730 Series Maintenance 6 24 Figure 6 8 Repackaging a 1730 Series instrument...
Page 185: ...Options...
Page 186: ......
Page 189: ...Replaceable Electrical Parts...
Page 190: ......
Page 222: ...1730 Series Replaceable Electrical Parts 8 32...
Page 223: ...Diagrams Circuit Board Illustrations...
Page 224: ......
Page 246: ......
Page 247: ...Replaceable Mechanical Parts...
Page 248: ......
Page 255: ...1730 Series Waveform Monitor FIG 1 EXPLODED VIEW A1 A3 A2 A3A1 A10 A11...
Page 256: ...1730 Series B070000 Up...
Page 257: ......
Page 258: ......