1730–Series Checks and Adjustments
5–19
j.
CHECK – that the amplitude of the color bar is within 0.95 to 1.05 V
from sync tip to the 100% peak white bar, as displayed on the test
oscilloscope. Turn OFF the 1730–Series LINE SELECT.
k. Disconnect all signal cables from the 1730–Series Waveform Monitor.
12. Check DC Restorer Operation
REQUIREMENT – Attenuation of 60 Hz input signal 20% or less. Blanking
level shift with APL change, less than 1 IRE (7 mV). Blanking level shift with
presence or absence of burst, less than 1 IRE (7 mV).
a. Connect a modulated 5–step linearity signal (with AC Bounce on) to the
CH–B INPUT, and teminate in 75
Ω.
. Display the signal with the 2H
SWEEP. Turn ON FAST DC REST. Position the blanking level of the
signal to the 0 IRE (0 V) line.
b. CHECK – that the blanking level does not move when the VARIABLE
gain is rotated.
c.
CHECK – that the blanking level moves when the DC REST is turned
off. Leave DC REST OFF. Turn off variable gain.
d. Connect the black burst signal to the EXT REF and terminate the
remaining side of the loop–through input.
e.
Connect the function generator output through a X10 (75
W
) Attenuator
to the CH–A INPUT.
f.
Select CH–A and EXT REF.
g. Set the function generator frequency to a 60 Hz (50 Hz PAL) sine wave.
Set the amplitude for a 100 IRE (700 mV PAL) 1730–Series display.
h. Turn ON SLOW DC REST.
i.
CHECK – that the display amplitude is 80 IRE (560 mV PAL) or
greater.
j.
Turn ON FAST DC REST.
k. CHECK – that the display amplitude is 10 IRE (70 mV PAL) or less.
l.
Remove the function generator output and replace it with the linearity
signal and terminate the remaining side of the loop–through input.
m. Select 2LINE SWEEP.
n. Turn ON SLOW DC REST.
Summary of Contents for 1730 Series
Page 4: ......
Page 12: ...Contents viii...
Page 17: ...Introduction and Specifications...
Page 18: ......
Page 32: ...1730 Series Introduction 1 14...
Page 33: ...Operating Instructions...
Page 34: ......
Page 62: ...1730 Series Operating Instructions 2 28...
Page 64: ...Service Safety Summary S 2 1730 Series B070000 Above...
Page 66: ......
Page 67: ...Installation...
Page 68: ......
Page 82: ...1730 Series Installation 3 14...
Page 83: ...Theory of Operation...
Page 84: ......
Page 115: ...Checks and Adjustments...
Page 116: ......
Page 159: ...Maintenance...
Page 160: ......
Page 180: ...1730 Series Maintenance 6 20 3 Remove the board by slipping it through the front panel opening...
Page 184: ...1730 Series Maintenance 6 24 Figure 6 8 Repackaging a 1730 Series instrument...
Page 185: ...Options...
Page 186: ......
Page 189: ...Replaceable Electrical Parts...
Page 190: ......
Page 222: ...1730 Series Replaceable Electrical Parts 8 32...
Page 223: ...Diagrams Circuit Board Illustrations...
Page 224: ......
Page 246: ......
Page 247: ...Replaceable Mechanical Parts...
Page 248: ......
Page 255: ...1730 Series Waveform Monitor FIG 1 EXPLODED VIEW A1 A3 A2 A3A1 A10 A11...
Page 256: ...1730 Series B070000 Up...
Page 257: ......
Page 258: ......