Checks and Adjustments
1705A Spectrum Monitor
5- 9
!
!
SG503
VARIABLE
AUTOTRANSFORMER
70 MHz
INPUT
1705A
Figure 5-3: Output of the SG503 connected directly to the 1705A 70 MHz INPUT
g.
Set the leveled sine wave generator Frequency to 45 MHz, Amplitude
Multiplier to .01, and Amplitude to 0.5 (--42 dBm).
h.
Set the 1705A SPAN/DIV to FULL, and the CENTER FREQUENCY to
read 45 MHz. Check to see that the marker is intensified.
i.
Set the 1705A SPAN/DIV to 1 MHz and check that the marker is on
screen.
j.
Position the marker under the readout caret with the CENTER FRE-
QUENCY control.
k. CHECK
-- that readout reads 45,
±
1 count.
l.
Set SPAN/DIV to 100 kHz.
m. CHECK
-- that the marker is on screen.
n.
Repeat steps h. through m., substituting 95 MHz.
o.
Leave SPAN/DIV at 100 kHz.
Summary of Contents for 1705A
Page 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Page 17: ...Preface xvi 1705A Spectrum Monitor ...
Page 18: ...Introduction ...
Page 19: ......
Page 38: ...Operating Instructions ...
Page 39: ......
Page 61: ......
Page 62: ...Installation ...
Page 63: ......
Page 72: ...Theory of Operation ...
Page 73: ......
Page 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Page 94: ...Checks and Adjustments ...
Page 95: ......
Page 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Page 124: ...Maintenance ...
Page 154: ...Options ...
Page 155: ......
Page 158: ...Replaceable Electrical Parts ...
Page 159: ......
Page 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Page 190: ...Diagrams Circuit Board Illustrations ...
Page 191: ......
Page 193: ......
Page 213: ......
Page 214: ...Replaceable Mechanical Parts ...
Page 215: ......
Page 223: ......