Checks and Adjustments
5- 18
1705A Spectrum Monitor
15.
Adjust Span/Div Magnifier Range (R38 and R37)
16.
Adjust Intensified Zone Range (R61 and R36)
17.
Adjust Vertical Gain (R176, R239, R240, and R232)
18.
Adjust 2 dB/Div Position (R140)
19.
Adjust 590 MHz Notch (C69)
Figure 5-7 shows the 1705A Spectrum Monitor internal adjustment locations.
PRELIMINARY SETUP
Connect the 1705A ac power cord to the variable autotransformer. Turn the
1705A power on. Set the front-panel controls to start this procedure as
shown in Table 5--2.
Table 5- 2: Preliminary Control Settings
POWER
ON
INTENSITY
Set to Preference
FOCUS
SCALE
VERTICAL POSITION
Set later
HORIZONTAL POSITION
INPUT
70 MHz
RESOLUTION
300 kHz
VIDEO
ON
SPAN/DIV
FULL
CENTER FREQUENCY
anywhere
SWEEP SPEED
y
A1 POWER SUPPLY BOARD
(Refer to Figure 5-7 for adjustment locations.)
1.
5 V
a.
Connect the DMM negative lead to TP1 (GND) and the positive lead to
W1 (+5V).
b.
ADJUST -- R99 (+5V ADJ) for +5.0 V
0.5V.
Long Form Procedure
Summary of Contents for 1705A
Page 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Page 17: ...Preface xvi 1705A Spectrum Monitor ...
Page 18: ...Introduction ...
Page 19: ......
Page 38: ...Operating Instructions ...
Page 39: ......
Page 61: ......
Page 62: ...Installation ...
Page 63: ......
Page 72: ...Theory of Operation ...
Page 73: ......
Page 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Page 94: ...Checks and Adjustments ...
Page 95: ......
Page 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Page 124: ...Maintenance ...
Page 154: ...Options ...
Page 155: ......
Page 158: ...Replaceable Electrical Parts ...
Page 159: ......
Page 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Page 190: ...Diagrams Circuit Board Illustrations ...
Page 191: ......
Page 193: ......
Page 213: ......
Page 214: ...Replaceable Mechanical Parts ...
Page 215: ......
Page 223: ......