Checks and Adjustments
5- 20
1705A Spectrum Monitor
3.
Adjust Geometry, Focus, and Astigmatism
Display the DAC test pattern by completing parts a. through c.
a.
Press both the 1705A SPAN/DIV buttons simultaneously to display the
Main Menu.
b.
Press either SPAN/DIV button to move the cursor to TEST, then press
the INPUT button to display the TEST menu.
c.
Press either SPAN/DIV button to move the cursor to DAC, then press
the INPUT button to display the DAC test pattern.
d.
Set the front-panel FOCUS control so that it is approximately at the
center of its rotation.
e.
ADJUST
-- R45 (GEOM) for 35V
1V at pin 1 of J3.
f.
ADJUST
-- R11 (CTR FOCUS) and R49 (ASTIG) for the most
clearly-defined DAC test pattern. See Figure 5-8.
Tek
REF
--10
--20
--30
--40
--50
--60
--70
L
O
G
+ f
Figure 5-8: DAC check waveforms used to check Focus, Astigmatism, Geometry,
and setting of the Trace Rotation (DAC test pattern may vary in size from instru-
ment to instrument)
4.
Adjust Trace Rotation
a.
ADJUST
-- the front-panel TRACE ROTATION potentiometer for a
level display, using a crt graticule line as a reference.
Summary of Contents for 1705A
Page 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Page 17: ...Preface xvi 1705A Spectrum Monitor ...
Page 18: ...Introduction ...
Page 19: ......
Page 38: ...Operating Instructions ...
Page 39: ......
Page 61: ......
Page 62: ...Installation ...
Page 63: ......
Page 72: ...Theory of Operation ...
Page 73: ......
Page 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Page 94: ...Checks and Adjustments ...
Page 95: ......
Page 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Page 124: ...Maintenance ...
Page 154: ...Options ...
Page 155: ......
Page 158: ...Replaceable Electrical Parts ...
Page 159: ......
Page 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Page 190: ...Diagrams Circuit Board Illustrations ...
Page 191: ......
Page 193: ......
Page 213: ......
Page 214: ...Replaceable Mechanical Parts ...
Page 215: ......
Page 223: ......