Checks and Adjustments
5- 12
1705A Spectrum Monitor
10. Check L--Band Linearity
REQUIREMENT
-- One marker per Division
±
1 minor Division.
a.
Connect the output of the leveled sine wave generator, through the comb
generator, to the 1705A L--BAND INPUT. See Figure 5-4.
SG503
Comb
Generator
VARIABLE
AUTOTRANSFORMER
L--BAND
INPUT
1705A
Figure 5-4: Equipment connections for L- Band checkout
b.
Set the leveled sine wave generator Frequency to 100 MHz, Amplitude
Multiplier to X1, and the Amplitude to 1.5 (8 dBm).
c.
Set the 1705A INPUT to L--BAND.
d.
Set the SPAN/DIV to FULL.
e.
CHECK
-- for 10 frequency markers. (Ignore any sweep start marker.)
f.
CHECK
– that each marker is within one minor Division of a major
graticule line when the 2
nd
marker is on the first graticule line in from
the left. See Figure 5-5.
Summary of Contents for 1705A
Page 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Page 17: ...Preface xvi 1705A Spectrum Monitor ...
Page 18: ...Introduction ...
Page 19: ......
Page 38: ...Operating Instructions ...
Page 39: ......
Page 61: ......
Page 62: ...Installation ...
Page 63: ......
Page 72: ...Theory of Operation ...
Page 73: ......
Page 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Page 94: ...Checks and Adjustments ...
Page 95: ......
Page 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Page 124: ...Maintenance ...
Page 154: ...Options ...
Page 155: ......
Page 158: ...Replaceable Electrical Parts ...
Page 159: ......
Page 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Page 190: ...Diagrams Circuit Board Illustrations ...
Page 191: ......
Page 193: ......
Page 213: ......
Page 214: ...Replaceable Mechanical Parts ...
Page 215: ......
Page 223: ......