Checks and Adjustments
1705A Spectrum Monitor
5- 13
ALIGN 2nd
MARKER HERE
DISREGARD
#1
GRATICULE
LINE
#2
GRATICULE
LINE
Figure 5-5: Aligning L- Band time markers with the graticule lines. The graticule
begins with 0 while the time markers are counted 1, 2, 3, etc. The first and last
divisions of sweep should be disregarded
11. Check L--Band SPAN/DIV and Readout
REQUIREMENTS:
FULL —
1 marker/Division
±
1 minor Division
10 MHz —
1 marker/5 Divisions
±
1 Division
1 MHz —
1 marker/5 Divisions
±
1 Division
Readout Accuracy —
±
20 MHz.
a.
Set CENTER FREQUENCY to 1400 MHz, and set SPAN/DIV to 10
MHz.
b.
Set the leveled sine wave generator Frequency to 50 MHz.
c.
Position the readout caret to center of screen horizontally.
d. CHECK
-- for 1 marker every 5 Divisions,
±
1 Division (adjust
CENTER FREQUENCY, if necessary).
e.
Repeat parts a. through d. with CENTER FREQUENCY at 950 MHz
and 1700 MHz.
f.
Set CENTER FREQUENCY to 1400 MHz.
g.
Set SPAN/DIV to 1 MHz.
Summary of Contents for 1705A
Page 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Page 17: ...Preface xvi 1705A Spectrum Monitor ...
Page 18: ...Introduction ...
Page 19: ......
Page 38: ...Operating Instructions ...
Page 39: ......
Page 61: ......
Page 62: ...Installation ...
Page 63: ......
Page 72: ...Theory of Operation ...
Page 73: ......
Page 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Page 94: ...Checks and Adjustments ...
Page 95: ......
Page 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Page 124: ...Maintenance ...
Page 154: ...Options ...
Page 155: ......
Page 158: ...Replaceable Electrical Parts ...
Page 159: ......
Page 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Page 190: ...Diagrams Circuit Board Illustrations ...
Page 191: ......
Page 193: ......
Page 213: ......
Page 214: ...Replaceable Mechanical Parts ...
Page 215: ......
Page 223: ......