Checks and Adjustments
5- 10
1705A Spectrum Monitor
6.
Check Resolution Filter
REQUIREMENT
-- 300 kHz Filter Resolution 200 -- 400 kHz. 10 kHz
Filter Resolution 20 kHz or less. Filter through gain matches
±
2 dB.
a.
Adjust the leveled sine wave generator Amplitude for a 5 Division signal
amplitude.
b. CHECK
-- that the marker width 0.6 Divisions down from the top of the
signal is 3 Divisions,
±
1 Division.
c.
Change resolution to 10 kHz.
d.
Set SWEEP SPEED fully counterclockwise.
e.
CHECK
-- that the marker width 0.6 Divisions down from the top of the
signal is
≤
2 minor Divisions.
f.
CHECK
-- that the 10 kHz marker amplitude matches the highest point
of the 300 kHz marker amplitude within 1 minor Division, when
switching between 300 KHZ and 10 KHZ RESOLUTION.
7.
Check 70 MHz Gain and Flatness
REQUIREMENT
-- Incremental Amplitude Accuracy
±
4 dB/70 MHz.
Flatness
±
1 minor Division from 70 MHz to either 50 or 90 MHz.
a.
Set SPAN/DIV to FULL.
b.
Set RESOLUTION to 300 kHz.
c.
Set SWEEP SPEED fully clockwise.
d.
Set the leveled sine wave generator Frequency to 70 MHz, the Ampli-
tude Multiplier to X1, and the Amplitude to 0.5 (--2 dBm).
e.
CHECK
-- for a marker amplitude change of 2 Divisions,
±
2 minor
Divisions when switching the leveled sine wave generator Amplitude
Multiplier between 0.01 and 0.1 (a 20 dB change).
f.
Set the leveled sine wave generator Amplitude Multiplier to 0.01 and
Amplitude to 0.5 (--42 dBm).
g.
Position the tip of the marker to exact crt center with the HORIZONTAL
and VERTICAL POSITION controls.
h.
Set the leveled sine wave generator Frequency to 50 MHz.
i.
CHECK
-- that the tip of the marker is on the crt center line,
±
1 minor
Division (vertically).
j.
Set the leveled sine wave generator Frequency to 90 MHz.
Summary of Contents for 1705A
Page 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Page 17: ...Preface xvi 1705A Spectrum Monitor ...
Page 18: ...Introduction ...
Page 19: ......
Page 38: ...Operating Instructions ...
Page 39: ......
Page 61: ......
Page 62: ...Installation ...
Page 63: ......
Page 72: ...Theory of Operation ...
Page 73: ......
Page 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Page 94: ...Checks and Adjustments ...
Page 95: ......
Page 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Page 124: ...Maintenance ...
Page 154: ...Options ...
Page 155: ......
Page 158: ...Replaceable Electrical Parts ...
Page 159: ......
Page 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Page 190: ...Diagrams Circuit Board Illustrations ...
Page 191: ......
Page 193: ......
Page 213: ......
Page 214: ...Replaceable Mechanical Parts ...
Page 215: ......
Page 223: ......