Introduction
1- 6
1705A Spectrum Monitor
Table 1- 4: Environmental Characteristics
Characteristic
Supplemental Information
Temperature
Non
-
Operating
Operating
--- 55˚C to +75˚C.
0˚C to +50˚C.
Altitude
Non
-
Operating
Operating
To 50,000 feet (15,000 meters).
To 15,000 feet (4,800 meters).
Vibration --- Operating
15 minutes each axis at 0.015”, frequency varied from 10--- 55--- 10 Hz in 1
-
minute
cycles with instrument secured to vibration platform. Ten minutes each axis at
any resonant point or at 55 Hz if no resonant point is found.
Shock --- Non
-
Operating
30 g’s, ½ sine, 11 ms duration, 3 shocks per surface (18 total).
Transportation
Qualified under NTSC Test Procedure 1A, Category II (30” drop).
Humidity
Will operate at 95% relative humidity for up to five days.
Table 1- 5: Physical Characteristics
Characteristic
Supplemental Information
Dimensions
Height
Width
Length
5 1/4 inches (133.4 mm).
8 1/2 inches (215.9 mm).
18 1/8 inches (460.4 mm).
Weight
Approximately 8.5 lbs (approximately 3.8 kg).
Table 1- 6: Certifications and Compliances
EC Declaration of Conformity --
EMC
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 50081-1 Emissions:
EN 55022
Class B Radiated and Conducted Emissions
EN 50082-1 Immunity:
IEC 801-2
Electrostatic Discharge Immunity
IEC 801-3
RF Electromagnetic Field Immunity
IEC 801-4
Electrical Fast Transient/Burst Immunity
High-quality shielded cables must be used to ensure compliance to the above listed standards.
This product complies when installed into any of the following Tektronix instrument enclosures:
1700F00 Standard Cabinet
1700F02 Portable Cabinet
1700F05 Rack Adapter
An increase of up to 20dB in the displayed noise floor may be observed if this instrument is
operated in electromagnetic fields of 3V/M or more, at frequencies of approximately 130, 250, 350,
or 490 MHz.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - FAX 781.665.0780 - TestEquipmentDepot.com
Summary of Contents for 1705A
Page 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Page 17: ...Preface xvi 1705A Spectrum Monitor ...
Page 18: ...Introduction ...
Page 19: ......
Page 38: ...Operating Instructions ...
Page 39: ......
Page 61: ......
Page 62: ...Installation ...
Page 63: ......
Page 72: ...Theory of Operation ...
Page 73: ......
Page 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Page 94: ...Checks and Adjustments ...
Page 95: ......
Page 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Page 124: ...Maintenance ...
Page 154: ...Options ...
Page 155: ......
Page 158: ...Replaceable Electrical Parts ...
Page 159: ......
Page 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Page 190: ...Diagrams Circuit Board Illustrations ...
Page 191: ......
Page 193: ......
Page 213: ......
Page 214: ...Replaceable Mechanical Parts ...
Page 215: ......
Page 223: ......