5. Operating the INFINITE M1000 PRO
68
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
5.7 Injectors
The
INFINITE M1000 PRO
can be optionally equipped with an injector module
consisting of one or two syringe pumps (XE-1000, Tecan Systems) located in a
separate box, which feed one or two injector needles (see figure below). Needles
are designed to inject liquid into any well of a 6-well to 384-well microplate.
Three different syringe volumes are available. The standard volume is 1000 µl,
additionally syringes with volumes of 500 µl and 2500 µl are offered.
Figure 34
: View of the injector module.
This label appears on the rear
panel of the injector module.
The
INFINITE M1000 PRO
can be equipped with one pump or two pumps
•
Pump A feeds injector needle A
•
Pump B feeds injector needle B
One Injector Option (one pump):
An
INFINITE M1000 PRO
equipped with one
pump allows injections in all wells of 6 to 384-well plates that comply to SBS
standard. All reactions requiring injection of only one liquid per well can be
performed with this option.
Two Injector Option (two pumps):
Several reactions, such as flash
luminescence reactions or dual reporter gene assays require the injection of two
independent
liquids into the same well, therefore Tecan Austria offers a two
injector option. By using pumps A and B, two independent liquids can be
dispensed into the same well of a 6-well to 384-well microplate.
WARNING
Switch off the instrument before plugging in or unplugging
the injector module.