3.
Installation
2011-09
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
25
3.3 Power
Requirements
The instrument is auto-sensing and it is therefore unnecessary to make any
changes to the voltage range. Check the voltage specifications on the rear panel
of the instrument and ensure that the voltage supplied to the instrument is correct
to this specification.
The voltage range is from 100–120 V and 220–240 V, 50/60 Hz.
If the voltage is not correct, please contact your distributor.
Connect the instrument only to an electricity supply system with protective earth.
STOP
Caution
Do not use the instrument if the voltage setting is not correct.
If the instrument is switched ON with the incorrect voltage
setting it will be damaged.
3.4
Switching the Instrument ON
STOP
Caution
Before the instrument is switched on for the first time after
installation, it should be left to stand for at least 3 hours, so there
is no possibility of condensation causing a short circuit.
•
Ensure the computer is switched OFF and the instrument's main power
switch in the back panel of the instrument is in the OFF position.
•
Connect the computer to the instrument only with the delivered USB interface
cable.
•
Insert the power cable into the main power socket (with protective earth
connection) in the back panel of the instrument.
•
All connected devices must be approved and listed as per IEC 60950-1
Information Technology Equipment – Safety or equivalent local standards.
•
Switch the instrument ON using the main power switch on the back panel of
the instrument.
WARNING
Switch off the instrument before plugging in or unplugging
the injector module.
STOP
Caution
When installing or uninstalling the instrument, ensure that the
instrument and the computer are both switched off and disconnected
from the main power supply before the USB interface cable or any other
cables are connected or removed.