7. Quality Control
112
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
7.4.4.3 Crosstalk ATP 384-well Plate:
Material:
Same plate/plate layout and reagent as described in chapter 7.4.4.2 Detection
Limit ATP 384-well Plate
Evaluation:
Calculate the crosstalk from measurement results of chapter 7.4.4.2 Detection
Limit ATP 384-well Plate for each well separately. The following wells are
charged together: A2 with A1, A2 with A3, B2 with B1, B2 with B3, C2 with C1,
C2 with C3, D2 with D1, and D2 with D3. The average of these 8 values must be
within the specified limits, see chapter 7.3 Acceptance Criteria (under
Luminescence Crosstalk).
)
(
100
*
)
(
%
B
ATP
B
B
CT
wellx
X
−
−
=
B
X
Blank wells A1 – D1 and A3 – D3, respectively
ATP
wellx
ATP wells A2 – D2, respectively
B