6. Instrument Specifications
94
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
6.8
“On the Fly” Measurements
“On the Fly” measurements are the fastest measurements possible using the
INFINITE M1000 PRO
. These measurements are performed with one flash only,
a lag time of 0, and an integration time shorter than 100 µs.
Measurement Technique
Measurement Time
96-well 384-well 1536-well
Absorbance
11 s
15 s
23 s
Fluorescence Intensity Top
17 s
24 s
36 s
Fluorescence Intensity Bottom
19 s
25 s
38 s
6.9 Injectors
Parameters Characteristics
Plate Types
6 to 384-well plates
Injector Syringe Volumes
500 µl, 1000 µl, 2500 µl
6.9.1 Injector
Performance
Injector Syringe Volume
500 µl
1000 µl
2500 µl
Accuracy at 10 µl
≤
10 %
≤
10 %
≤
20 %
Accuracy at 100 µl
≤
2 %
≤
2 %
≤
5 %
Accuracy at 450 µl
≤
0.7 %
≤
0.7 %
≤
2 %
Precision at 10 µl
≤
10 %
≤
10 %
≤
20 %
Precision at 100 µl
≤
2 %
≤
2 %
≤
5 %
Precision at 450 µl
≤
0.7 %
≤
0.7 %
≤
2 %
Injector specifications are valid for 96-well plates only.