5. Operating the INFINITE M1000 PRO
2011-09
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
77
5.7.5
Injector Modes and Settings (i-control)
When using the injector, two modes are available:
Dispense
: The dispense mode allows liquid to be dispensed plate-wise into the
selected wells
Injection
: This mode must be used in combination with a measurement strip.
The injection is performed in a well-wise mode.
Dispense Mode
The dispense settings can be adjusted via the software:
Figure 38
Dispense
Select Injector
: Injector A and/or Injector B can be selected.
Speed
: The injection speed is selectable from 100 – 300
μ
l/sec for each injector.
Select ‘
Refill speed
’ from 100 – 300
μ
l/sec. for each injector or select ‘
Refill
Speed equal to Dispense Speed
’.
Select refill mode ‘
Standard
’, if refill should be performed when syringe is empty.
The
Refill volume
can be defined for each injector separately and allows
economical handling of solutions.
Select ‘
Refill for every dispense
’, if refill should be performed for every
dispense step.
The option
‘Read time like dispense time’
aligns the measurement time to the
dispensing time in order to achieve same reaction times for each measured well.