3. Installation
22
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
4. Lift the instrument out of the carton and place it in the selected location. Take
care when lifting the instrument and ensure that it is held on both sides.
5. Visually inspect the instrument for loose, bent or broken parts.
Report any damage immediately.
6. Compare the serial number on the rear panel of the instrument with the serial
number on the packing slip.
Report any discrepancy immediately.
7. Check the instrument accessories against the packing list.
8. Save packing materials and transport locks (see next section) for further
transportation purposes.
WARNING
The fully equipped INFINITE M1000 PRO is a precision instrument
and weighs approximately 29.5 kg. At least two people must
carefully lift the instrument from the box.
STOP
Caution
The maximum load for the INFINITE M1000 PRO cover is 20 kg;
however, the load must be distributed evenly across the entire
surface of the cover.
STOP
Caution
The maximum load for the INFINITE M1000 PRO plate transport is
300 g. Overloading the plate carrier can cause instrument damage
which may require service.
Plate carrier testing and wavelength calibration should be done
annually with the MultiCheck-Plus Test Plate to ensure the
optimal performance of the INFINITE M1000 PRO.
STOP
Caution
Allow at least 10 cm distance between the back of the instrument
and the wall or any other equipment. Do not cover instrument
while it is in operation.