BRAKE / NEUTRAL / DRIVE POTENTIOMETER REMOVAL AND REPLACEMENT - QDF27-2024
Tools Required:
• Diagonal Pliers
Procedure:
1. Plug bed into wall outlet.
2. Raise the bed lift and patient’s left siderails all the way up.
3. Remove the base center covers (foot, center, and head).
4. Remove the patient’s left base frame cover.
5. Cut the two wire ties securing the potentiometer to the potentiometer bracket; then cut the two wire ties securing
the potentiometer wires to the frame.
6. Remove the electrical tape holding the quick connection together.
7. Reverse the above procedures for installation of the new potentiometer.
Note: Make sure to install new wire ties in the same location.
8. Recalibrate the bed (refer to Bed Calibration procedures on
page 25
).
9. Test all bed functionality prior to putting the bed back into service.