Setup
To unpack your product, see the unpacking instructions that are attached to the product inside of the shipping crate.
WARNING
• Always allow the product to reach room temperature before conducting any setup or testing functional operations
to prevent permanent product damage.
• Always operate the product when all operators are clear of the mechanisms.
• Always properly handle the power cord to avoid the risk of entanglement, damage to the power cord, or potential
shock hazards. If the power cord is damaged, immediately remove the product from service and contact the
appropriate maintenance personnel.
• Do not attach the power cord to any moving parts of InTouch.
• Always unplug the power cord, turn the battery switch to the
OFF (O)
position, press the N/Brake Off button, and
call maintenance if unanticipated motion occurs.
CAUTION
Always plug the product directly into a properly grounded, three-prong receptacle. You can only achieve grounding
reliability when you use a hospital-grade receptacle. This product is equipped with a hospital-grade plug for protection
against electric shock hazard.
To turn on the product:
1. Plug the power cord into a properly grounded, hospital-grade wall outlet.
2. Turn the battery switch to the
ON (|)
position.
Before placing the product into service, make sure that these components are working properly:
1. Visually inspect the product for any signs of shipping damage.
2. Flip down and depress the manual brake pedal and make sure that the neutral, drive, and brake functions of the
manual brake pedal hold.
3. Press BRAKE on each control panel and make sure that the neutral, drive, and brake functions of the electronic
brake hold.
4. Test the Zoom drive system (Model 2141).
5. Raise and lower the siderails to make sure that they move smoothly and lock securely in the full up position.
6. Press each button on the head end control panel (optional), motion control panel, brake control panel, footboard
control panel, and patient control pendant (optional) to make sure that each function operates properly.
7. Make sure that the footboard operates properly.
8. Make sure that the scale system operates properly.
9. Make sure that the bed exit system operates properly.
10. Make sure that the CPR release pedal operates properly.
11. Make sure that the support surface operates properly (optional).
12. Make sure that optional accessories are installed and operate as described.
WARNING
Always unplug the product power cord from the wall when using oxygen administering equipment. Possible fire hazard
exists when this product is used with oxygen administering equipment other than nasal, mask type, or half bed-length
tent type.