Maintenance Menu Guide
INTOUCH CONFIGURATION SCREEN (CONTINUED)
A. Bed Calibration (Continued)
Figure 25: Calibration Complete
31. When the calibration procedure is completed, Figure 25 will be displayed as shown below.
32. Press the Close button to exit the Calibration Procedure Menu.
33. Level the litter flat using the CPR pedal. Verify all display readings are zero degrees while the bed is at
33” (84 cm) (highest height). Lower the bed to 16” (41 cm) (lowest height) and verify the display readings
maintain a constant reading of zero degrees with no fluctuations. When the bed reaches it lowest height at
16”, verify the display readings are still at zero degrees.
Note: If readings are not all zero, you will need to recalibrate the bed by repeating steps 1-30.