Return To Table of Contents
60
2141-209-002 REV H
www.stryker.com
Error Codes
Touch Screen Error Messages Name Definition
Scale Chip Failure
Control board is bad (Replace Control Board).
Tilt base over range
Verify if tilt sensor is still in place. Inspect for a damaged or improperly
assembled tilt sensor or recalibrate bed.
Tilt Error Base
Angle sensor is damaged (Replace sensor).
Tilt Error Foot
Angle sensor is damaged (Replace sensor).
Tilt Error Gatch
Angle sensor is damaged (Replace sensor).
Tilt Error Head
Angle sensor is damaged (Replace sensor).
Tilt Error HiLo Foot
Angle sensor is damaged (Replace sensor).
Tilt Error Trend
Angle sensor is damaged (Replace sensor).
Tilt foot over range
Verify if tilt sensor is still in place. Inspect for a damaged or improperly
assembled tilt sensor or recalibrate bed.
Tilt gatch over range
Verify if tilt sensor is still in place. Inspect for a damaged or improperly
assembled tilt sensor or recalibrate bed.
Tilt head over range
Verify if tilt sensor is still in place. Inspect for a damaged or improperly
assembled tilt sensor or recalibrate bed.
Tilt hiLo foot over range
Verify if tilt sensor is still in place. Inspect for a damaged or improperly
assembled tilt sensor or recalibrate bed.
Tilt trend over range
Verify if tilt sensor is still in place. Inspect for a damaged or improperly
assembled tilt sensor or recalibrate bed.
Zoom® time out for switch
Verify wiring to Zoom® motor.
Wi-Fi / IR Communication Error
Definition
Wireless Adapter Failure
Unable to detect or communicate with Wi-Fi radio. Check assembly or
cabling of Wi-Fi card in footboard or replace Wi-Fi card or footboard.
DHCP Error: Unable to obtain IP
Address
Bed is not able to connect to the wireless network. Check wireless network
settings and Wi-Fi configuration of bed.
Unable to connect to network
Bed is not able to connect to the wireless network. Check wireless network
settings and Wi-Fi configuration of bed.
Internal Communication Error
Internal software failure. Software corruption may have occurred. Re-upload
software or replace footboard.
Wi-Fi Settings corrupted
Reconfigure Wi-Fi settings in bed.
Bed Serial Number not available
Configure bed serial number.
Left IR Module not detected
Unable to detect IR module. Check cabling or replace the IR module.
Right IR Module not detected
Unable to detect IR module. Check cabling or replace the IR module.
IR Interface Board not detected
Unable to detect the Room interface board. Check cabling or replace the
Room interface board.
IR Crosstalk detected
Crosstalk detected with iBed Locator Wall module. Verify placement of IR
modules on bed or wall modules.
Isolibrium Error Code
Definition
0100
Sensor Board communication failure
0101
Sensor Board EEPROM corruption
0102
Sensor Board not calibrated
0200
GUI CANopen Heartbeat Lost
0400
EEPROM corruption
0500
Ambient temperature sensor failure
0501
Ambient temperature out of range
0600
Pressure sensor failure
0601
Seat Left pressure sensor failure