INTOUCH CONFIGURATION SCREEN (CONTINUED)
F. Serial Number Configuration
To access the Serial Number screen:
1. Press and hold the
HOB 30
button and the
Vascular Position
button at the same time for 5 seconds then release
both buttons. The Bed Configuration - Serial Number screen (Figure 36) will be displayed.
2. Enter or confirm the serial number of the bed (Figure 36).
a. To enter the serial number
:
i.
Press the pencil button located to the right of the serial
number field. The Edit screen will be displayed.
ii. Enter the serial number of the bed in the serial number
field, then press the
OK
button. You will be returned to
the serial number main screen as shown in Figure 36.
iii. Press the
Continue
button and proceed to step 3.
b. To confirm the serial number
:
i.
Review the serial number displayed in the serial number
field.
ii. If serial number is correct, press the
Back
button and
proceed to step 6.
iii. If the serial number is incorrect, press the pencil
icon
button to open the Edit screen then enter the correct
serial number.
iv. Press the
OK
button and you will be returned to the
serial number main screen.
v. Press the
Continue
button and proceed to step 3.
3. After pressing the
Continue
button, the serial number will be
saved and the serial number confirmation screen (Figure 37) will
be displayed. Press the
OK
button.
4. After pressing the
OK
button, a cycle power screen (Figure 38)
will be displayed stating to cycle power on the bed.
5. To cycle power on the bed, turn the battery disconnect switch to
OFF
(O) then unplug the power cord from the wall outlet. Plug
the power cord back into the wall outlet then turn the battery
disconnect switch back
ON
(|).
6. Test bed functionality prior to returning the bed into service.
Figure 37: Serial Number
Confirmation Screen
Figure 38: Cycle Power Screen
Figure 36: Bed Configuration Serial Number Screen
Pencil icon
Press to edit