Return To Table of Contents
96
2141-209-002 REV H
www.stryker.com
Service Information
TOUCH SCREEN REMOVAL AND REPLACEMENT - (FOOTBOARD)
Tools Required:
• Phillips Screwdriver #2
• Stubby Phillips Screwdriver #2
• Small Slotted Screwdriver
• Utility Knife
• ESD System
Procedure:
Note: ESD protection is required for this procedure.
1. Remove the footboard assembly and place onto a nearby work surface, face down.
2. Using a small slotted screwdriver or a utility knife, remove the rear footboard label.
Note: This label will need to be replaced with a new label (QDF27-2756).
3. Using a Phillips screwdriver, remove the six screws securing the footboard display panel to the footboard.
4. Pivot the display panel up.
5. Using a stubby Phillips screwdriver, remove the two screws from the bottom of the display panel.
6. Hold the display panel and footboard assembly together and flip the footboard assembly over, with the display
panel facing up.
7. Carefully remove the display panel from the footboard assembly and rotate over to display the circuit boards.
8. Disconnect the two ribbon cables from J3 and J7 by pulling outward on the locking tabs, then pulling the ribbon
cable.
9. Disconnect the menu board cable from connector J6.
10. Using a Phillips screwdriver, remove the two screws that secure the Wi-Fi board to the LCD interface board.
11. Using a Phillips screwdriver, remove the two screws that secure the LCD interface board to the display housing.
12. Remove the two spacer shims between the LCD interface board and the display housing.
13. Carefully remove the LCD interface board from the display.
Note: The components are secured together with double-sided foam tape. Slight upward force will be
necessary to separate.
14. Using a small slotted screwdriver, separate the locking tabs from the display and remove the display.
15. Using a small slotted screwdriver or utility knife, remove the foot control overlay label.
Note: This overlay label will need to be replaced with a new overlay label (QDF27-2188-ENG) and patent
pending label (QE71-1094-ENG).
16. Carefully separate the touch screen from the display housing.
17. Reverse the procedure to install the new touch screen.
18. Recalibrate the touch screen, by referencing the Touch Screen Calibration procedure on
page 25
.
19. Test all bed functionality before returning the bed to service.
Touch
Screen
Rear Footboard Back Label
Screws
Display