Stryker InTouch FL27 series Maintenance Manual Download Page 1

Maintenance Manual

In

Touch

®

 Critical Care Bed

REF

 FL27 (2131/2141)

2018/03  H.0

2141-209-002 REV H

www.stryker.com

Version 4.0 with Wi-Fi / 

Isolibrium™ (2971) support surface 

Summary of Contents for InTouch FL27 series

Page 1: ...Maintenance Manual InTouch Critical Care Bed REF FL27 2131 2141 2018 03 H 0 2141 209 002 REV H www stryker com Version 4 0 with Wi Fi Isolibrium 2971 support surface ...

Page 2: ......

Page 3: ... Type B Applied Part 87VL Medical Equipment Classified by Underwriters Laboratories Inc With Respect to Electric Shock Fire and Mechanical Hazards Only in Accordance with ANSI AAMI ES60601 1 2005 and CAN CSA C22 2 No 60601 1 08 In accordance with European Directive 2012 19 EU on Waste Electrical and Electronic Equipment this symbol indicates that the product must not be disposed of as unsorted mun...

Page 4: ...Contents 4 2141 209 002 REV H www stryker com Symbols iBed Locator is connected iBed Locator is not connected Wireless Network is connected Wireless Network is not connected Support surface call maintenance Serial number ...

Page 5: ...the PositionPRO support surface optional 23 Installing the Isolibrium support surface optional 23 Installing iBed Wireless 120V North American only optional 23 Maintenance Menu Guide 24 InTouch Configuration Screen 25 A Bed Calibration 25 B Full Diagnostic 36 C Touch Screen Calibration 37 D Optional Wi Fi Configuration 38 E Bed Options Configuration 39 F Serial Number Configuration 40 Accessing th...

Page 6: ...moval and Replacement Footboard 97 Brake Neutral Drive Potentiometer Removal and Replacement QDF27 2024 98 Battery Removal and Replacement Litter 99 Fowler Angle Sensor Removal and Replacement Litter 100 Gatch Angle Sensor Removal and Replacement Litter 101 Foot Angle Sensor Removal and Replacement Litter 102 Base Angle Sensor Removal and Replacement Base 103 Trend Angle Sensor Removal and Replace...

Page 7: ...y Head End with Speaker with NC with iAudio with iBed Wireless 189 Siderail Mounting Assembly Foot End 193 Standard Siderail Assembly Foot End Right 194 Standard Siderail Assembly Foot End Left 197 Standard Siderail Assembly without iBed 200 Optional Siderail Assembly with iBed 201 Headboard and Footboard Assembly 202 Standard Footboard Assembly 203 Footboard Assembly without iBed 205 Footboard As...

Page 8: ... 2 in FA64217 XXX 230 Traction Sleeve Assembly 8 in x 3 4 in FA64218 XXX 231 Traction Sleeve Assembly 6 1 2 in x 3 4 in FA64219 XXX 232 Wall Saver Cable Assembly FA64208 233 X ray Cassette Holder Assembly FA64205 XXX 234 Tube Support 236 Recycling Passports 237 Warranty 249 Limited Warranty 249 Warranty exclusion and damage limitations 249 To Obtain Parts and Service 249 Return Authorization 249 D...

Page 9: ...ential serious adverse reactions and safety hazards CAUTION Alerts the reader of a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property This includes special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a resu...

Page 10: ...st in ventilator associated pneumonia VAP prevention When the Chaperone bed exit system is active it monitors a chosen zone and alerts the operator of a deliberate or non deliberate bed exit The bed has 39 prerecorded clinical phrases in 24 languages and a sound feature that offers various environmental and musical selections INTENDED USE INTOUCH CRITICAL CARE BED InTouch is intended for use by pa...

Page 11: ... not intended to provide automated treatment decisions or as a substitute for professional healthcare judgment The iBed Wireless with iBed Awareness System is not a replacement or substitute for vital signs monitoring or alert equipment All patient medical diagnosis and treatment are to be performed under direct supervision and oversight of an appropriate health care professional EXPECTED SERVICE ...

Page 12: ...40 50 5 Gatch Section Length Width Angle 18 in 34 in to 35 in 0 to 15 3 45 7 cm 86 4 cm 88 9 cm 0 to 15 3 Cardiac Chair Standard Cardiac Chair Position Enhanced Cardiac Chair Position Head 65 Seat 17 Foot 30 Trend 3 Head 70 Seat 19 Foot 47 Trend 3 Bed Lift System Height high to top of litter 33 in 83 8 cm Height low to top of litter 16 in 40 6 cm Trendelenburg Reverse Trendelenburg 12 2 Bed Lift T...

Page 13: ... CHAPv2 802 1x PEAP MS CHAPv2 Without certificate Supports IPv4 and DHCPv4 Battery 12 V 17 6 Ahr Sealed Lead Acid Battery Part Number QDF9188 3 0 V 220mAh Lithium Battery Size 20mm Varta Int CR2032 Coin Cell PC Battery Holder Size 20mm MDP Int BA2032 Note For Isolibrium specifications see the Isolibrium operations maintenance manual Recommended Support Surface Size 35 x 84 x 6 88 9 cm x 213 4 cm x...

Page 14: ... 50 F 10 C 104 F 40 C 40 F 40 C 158 F 70 C Relative Humidity Non Condensing 30 75 10 95 Atmospheric Pressure 700 hPa 1060 hPa 500 hPa 1060 hPa Scale Accuracy 64 F 18 C 77 F 25 C Specifications listed are approximate and may vary slightly from product to product or by power supply fluctuations Stryker reserves the right to change specifications without notice Introduction ...

Page 15: ...rail M Mattress retainer C Caster N Nurse call inside siderail optional D CPR release pedal O Motion control panel outside siderail E Foley bag hooks P Pendant holder F Footboard Q Roller bumpers G Footboard control panel R Siderail release levers H Headboard S Siderails I Head end control panel optional T Support surface optional J InTouch touch screen U Zoom drive system model 2141 optional K Is...

Page 16: ...anual online see https techweb stryker com Have the serial number A of your Stryker product available when calling Stryker Customer Service or Technical Support Include the serial number in all written communication SERIAL NUMBER LOCATION You can find the serial number plate behind the patient right siderail near the foot end of the product A SPECIFICATION LABEL LOCATION You can find the specifica...

Page 17: ...footboard with an iBed Wireless compatible product You will lose iBed Wireless functionality if you use an older version of the footboard You must verify all iBed Wireless functionality after installation Verify that the iBed locator communicates the product s positions and that iBed Wireless communicates with the wireless access point Failure to do may result in the loss of information or the tra...

Page 18: ...tending medical personnel to determine the degree of restraint necessary to ensure a patient s safety Failure to use the siderails properly could result in serious patient injury Always keep the siderails outside of the oxygen tent The scale system is intended to assist in the monitoring of the patient s weight variation Under no circumstances should its reading be used as sole reference for medic...

Page 19: ...cause the product to flip Do not allow the optional line management clip to interfere with a mechanical or electronic mechanism of the product Do not pinch tubes inside the clip Do not clean the clip with a liquid solution Always adjust the scale or bed exit system if an option is added while the scale or bed exit system is armed Do not place objects that exceed 40 lb 18 kg onto the optional monit...

Page 20: ... a mattress or support surface thicker than six in 15 4 cm Do not allow sharp objects to come into contact with the support surface that could puncture tear or cut the cover Make sure that you set the desired product parameters before enabling iBed Awareness Do not use accessories that cover the footboard and outside siderail LED light bars Do not turn off the iBed Awareness alarm You will lose ac...

Page 21: ...led Always use adequate static protection when servicing the electronic systems of the product All service personnel must use static protection whenever they are touching wires Static Antistatic Protection Equipment Includes 1 antistatic wrist strap 1 grounding plug 1 test lead with a banana plug on one end and an alligator clip on the other end Static Protection Procedure 1 Unplug the power cord ...

Page 22: ... wall outlet 2 Turn the battery switch to the ON position Before placing the product into service make sure that these components are working properly 1 Visually inspect the product for any signs of shipping damage 2 Flip down and depress the manual brake pedal and make sure that the neutral drive and brake functions of the manual brake pedal hold 3 Press BRAKE on each control panel and make sure ...

Page 23: ...s access point Failure to do may result in the loss of information or the transmission of incorrect information You must install the iBed Locators more than 71 apart from one another in the same room such as in a semi private room with more than one product Failure to do may result in the product transmitting information to the incorrect iBed Locator Do not use iBed Wireless to replace the existin...

Page 24: ...ences between the mechanical parts of the bed Mechanical damage could occur without supervision Only qualified personnel should perform the calibration Note Verify that the bed is on a level surface which does not have any slopes or inclines prior to entering into the calibration mode 1 Unseat and reseat the footboard and wait until the main control screen is displayed 2 Push and hold the Main Men...

Page 25: ...s NEXT to continue with calibration The following items A B and C are configuration buttons available and displayed on the Configuration screen Figure 3 A Bed Calibration B Full Diagnostic see page 36 C Touch Screen Calibration see page 37 D Optional Wi Fi Configuration see page 38 The following items D and E are configuration options available but are not displayed on the Configuration screen To ...

Page 26: ...y the Foot Up Fowler Down and Gatch Down buttons as shown in Figure 6 7 Lower the bed to the lowest position Press the Bed Height Down Head Lift Down and Reverse Trendelenburg Foot Lift Down buttons simultaneously as shown in Figure 7 Figure 5 Bed Calibration Step 1 of 6 Figure 6 HE Siderail Control Panel Right Side Shown Fowler Down QDF27 2237 Gatch Down Foot Up Figure 7 HE Siderail Control Panel...

Page 27: ...d Calibration Continued 8 Press the Next button when done The Do Not Touch Bed screen will appear as shown in Figure 8 9 When step 1 of the calibration procedure has completed step 2 of the calibration procedure will begin and Figure 9 will be displayed Figure 8 Bed Calibration Cancel Do Not Touch Bed Figure 9 Bed Calibration Step 2 ...

Page 28: ...el inclinometer on the bottom of the litter below the mattress retainer refer to Figure 11 Using the digital level inclinometer verify the bed is level 0 0 0 1 Note Cycle power on the digital level inclinometer prior to placing it on the bottom of the litter frame and do not zero calibrate the digital level inclinometer 12 Press the Next button when done The Do Not Touch Bed screen will appear as ...

Page 29: ...ep 3 of the calibration procedure will begin and Figure 13 as shown below will appear on the screen 14 Place the bed at 12 degrees Trendelenburg by pressing the Trendelenburg Foot Lift Up button as shown in Figure14 Verify 12degrees 0 1 withtheinclinometer you previously placed on the litter frame in step 11 Figure 14 Head Siderail Control Panel Figure 13 Bed Calibration Step 3 QDF27 2237 Trendele...

Page 30: ...15 Press the Next button when done 16 The Do Not Touch Bed screen will appear as shown in Figure 15 17 When step 3 of the calibration procedure has completed step 4 of the calibration procedure will begin and Figure 16 will be displayed as shown below Figure 15 Do Not Touch Bed Screen Bed Calibration Cancel Do Not Touch Bed Figure 16 Bed Calibration Step 4 ...

Page 31: ...r reads zero degrees 19 Place the bed at 12 degrees Reverse Trendelenburg by pressing the Bed Height Up Head Lift Up button as shown in Figure 17 below Verify 12 degrees 0 1 with the inclinometer you previously placed on the bottom of the litter frame in step 11 20 Press the Next button when done 21 The Do Not Touch Bed screen will appear as shown in Figure 18 Figure 17 Head Siderail Control Panel...

Page 32: ...tep 5 of the calibration procedure will begin and Figure 19 will appear on the screen as shown below 23 Place the bed at the highest height by pressing simultaneously the Trendelenburg Foot Lift Up button and Bed Height Up Head Lift Up button as shown in Figure 20 Figure 20 Head Siderail Control Panel Right Side Shown Bed Height Up Head Lift Up QDF27 2237 Trendelenburg Foot Lift Up Figure 19 Bed C...

Page 33: ...until the Gatch reaches its highest height lastly press the Foot Up button until the foot section reaches its highest height refer to Figure 21 below The foot section should be at a flat position 25 Press the Next button when done 26 The Do Not Touch Bed screen will appear as shown in Figure 22 Gatch Up Fowler Up Figure 21 Head Siderail Control Panel Right Side Shown QDF27 2237 Foot Up Figure 22 D...

Page 34: ... Foot Lift Up button and Bed Height Up Head Lift Up button see Figure 24 29 Place the Fowler and Gatch section of the bed at the highest height by first pressing the Fowler Up button then the Gatch Up button See Figure 24 30 Place the foot at the lowest position by pushing the Foot Down button until limit is met see Figure 24 CAUTION The Gatch must be raised to the highest height prior to running ...

Page 35: ... to exit the Calibration Procedure Menu 33 Level the litter flat using the CPR pedal Verify all display readings are zero degrees while the bed is at 33 84 cm highest height Lower the bed to 16 41 cm lowest height and verify the display readings maintain a constant reading of zero degrees with no fluctuations When the bed reaches it lowest height at 16 verify the display readings are still at zero...

Page 36: ...hen a function button is pushed Note This requires assistance to press the buttons on the head siderails or at the head end control e BUTTONS PRESSED Figure 27e Provides information on when the CPU has detected a button being pressed Note This requires assistance to press the buttons on the head siderails or at the head end control f SIGNAL VALUES Figure 27f Provides information on CPU voltages lo...

Page 37: ...y activate 2 Carefully press and briefly hold a stylus or your finger on the center of the cross hair shown on the screen 3 Repeat as the target moves around the screen There will be five different locations to press center lower right upper right upper left and lower left 4 When the last cross hair has been touched Figure 29 displays New calibration settings have been measured 5 Tap the screen to...

Page 38: ...figure wireless settings for the bed 2 Tap Wi Fi OFF to toggle the wireless network connection on 3 Tap the appropriate tab and then tap the pencil icon to configure the network settings Figure 30 Figure 33 4 Tap Save to save the network settings 5 After saving the configured network settings a power cycle notification appears 6 Unplug the power cord from the wall outlet 7 Turn the battery switch ...

Page 39: ...ed iAudio iDoc and Wi Fi All of the options selected will turn green 3 After pressing Continue in step 2 a confirmation screen as shown in Figure 35 will be displayed verifying the options you selected a If the options on the confirmation screen do not match the options on the bed press the Back button b If the options on the confirmation screen match the options on the bed press the OK button 4 A...

Page 40: ...layed in the serial number field ii If serial number is correct press the Back button and proceed to step 6 iii If the serial number is incorrect press the pencil icon button to open the Edit screen then enter the correct serial number iv Press the OK button and you will be returned to the serial number main screen v Press the Continue button and proceed to step 3 3 After pressing the Continue but...

Page 41: ... malfunction The entry to this menu is through a key combination in the Support Surface menu To access the diagnostic menu 1 Unplug the InTouch bed and turn the battery switch to off and wait one minute 2 Turn the battery switch to on and plug the InTouch bed into a wall outlet 3 Tap Support Surfaces 4 Enter the following key combination within 5 minutes of the support surface connection Figure 40...

Page 42: ...M ERROR HISTORY To enter the Error History screen tap the Error History tab on the Diagnostic Screen Figure 43 The error list is ordered from the oldest entry to the newest with a maximum of 10 errors To clear the Error History screen tap Clear Error History on the Diagnostic screen Figure 43 Notes Always tap Clear Error History after troubleshooting is complete If you clear the error history whil...

Page 43: ...he support surface The screen displays the actual and goal pressure values For other non pressure signals only the actual values are displayed To enter the Advanced Signal Values screen tap Advanced Signal Values on the Diagnostics screen Figure 44 To exit the Advanced Signal Values screen tap Back to return to the Diagnostics screen Figure 45 Figure 44 Diagnostic Main Screen Sample Figure 45 Adva...

Page 44: ...es The results of the test will not be displayed until all tests have been completed No results will be displayed if the tests are canceled Other diagnostic functions are disabled while the diagnostic test is running The diagnostic test runs seven tests at one time on the support surface and take approximately 45 minutes to complete They include the following Seat Left Seat Right Back Left Back Ri...

Page 45: ...rogress To cancel diagnostics tap stop on the lower portion of the Diagnostics screen Figure 47 When the test is complete the results of each test section is indicated by a red cross when there is a failure or a green check mark when successful Figure 48 Note When a failure is first detected all following pressure tests will be marked as a failure Figure 48 Diagnostic Test Result Sample Figure 47 ...

Page 46: ...less than 1 part bleach to 100 parts water Hand wash all surfaces of the product with warm water and mild detergent Dry thoroughly Avoid oversaturation and make sure the product does not stay wet longer than the chemical manufacturer s guidelines for proper disinfecting CAUTION Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning Some cleaning...

Page 47: ...derail panels for cracks or splits All covers are not damaged and are not producing sharp edges Night light operating properly CPR release operating properly All siderail motion functionality Siderails move latch and stow properly Siderail switches working properly iBed Awareness option iBed Awareness light bars on footboard and siderails are working properly iBed Awareness option All functions on...

Page 48: ... intact and have at least two links touching the floor Enclosure is free from wear tear stresses and mechanical damage No rust or corrosion of parts Labels for legibility proper adherence and integrity iBed Wireless Module and IR Module are intact and footboard icons are displaying iBed Wireless option Serial Number Completed by _______________________________________ Date _________________ _____ ...

Page 49: ...ke Control QDF27 1097 Board Footboard Function Section LED s QDF75 0010 Board Siderail Outside Bed Motion QDF27 1099 Board Wi Fi 27 2775 Fuse 8Amp Ceramic QDF2120 Fuse 10Amp Main Power QDF8078 Fuse 25Amp Cartridge QDF2119 Load Cell QDF27 1372 Motor Drive 2141 Zoom Only QDF27 1445 Motor Drive Actuator 2141 Zoom Only 27 2593 Position Sensor Brake Brake Off Drive QDF27 2024 Power Cord 27 2782 Speaker...

Page 50: ...del both head end casters RD27 1970 Caster Steer 2131 Model Only both foot end casters RD27 1971 Headboard Assembly 27 2583K Footboard Assembly without iBed 27 2285K Footboard Assembly with iBed and iAudio 27 2824K Footboard Assembly with iBed and Wi Fi 27 2827K Footboard Assembly with iBed without iAudio 27 2826K Quick Reference Replacement Parts ...

Page 51: ... 3 Verify there is power at the transformer connection J11 on the CPU Power board at the foot end A Check for 24VAC at J11 between the blue and red wires I If 24VAC is present go to step 4 II If 24VAC is not present check the 25 Amp fuse in the fuse holder on the red wire from the transformer If bad replace the fuse III If 24VAC is not present check the power cable quick connection going to the tr...

Page 52: ...n 1 red lead to pin 2 a If voltage is present replace the motor b If voltage is not present replace the CPU Power board No Bed Down Motion FOOT 1 Put the bed into the Bed Calibration menu A Using one of the head siderails push the reverse trend button I If the foot lift motor runs down recalibrate the bed refer to the Bed Calibration procedures located on page 25 a If recalibration does not work r...

Page 53: ...or runs down recalibrate the bed refer to the Bed Calibration procedures located on page 25 a If recalibration does not work replace the fowler angle sensor and recalibrate II If the fowler motor does not run down check for 24VDC at connector J5 while pressing the fowler down button Red lead to pin 1 black lead to pin 2 a If voltage is present replace the motor b If voltage is not present replace ...

Page 54: ... recalibration does not work replace the foot angle sensor and recalibrate II If the foot motor does not run up check for 24VDC at connector J1 while pressing the foot up button Red lead to pin 1 black lead to pin 2 a If voltage is present replace the motor b If voltage is not present replace the CPU Power board No Foot Down Motion 1 2 Ensure there is no debris in the foot end traction sockets If ...

Page 55: ...ate b If replacement of the trend angle sensor did not resolve the problem replace the CPU Power board No Cardiac Chair Motion 1 Check the touch screen fowler and foot section angle display for accuracy of the angle of the fowler and foot section I If not accurate recalibrate the bed refer to the Bed Calibration procedures located on page 25 a If recalibration does not work replace the fowler or f...

Page 56: ...edures located on page 25 b If replacement of the fowler angle sensor did not resolve the problem replace the CPU Power board No Electric Brake Motion 1 Verify the Brake Not Set LED is flashing and the Brake Set LED is OFF I If the Brake Set LED is ON check manual brake position a If manual brake pedal is in the brake position the bed should not move and is okay go to step 2 b If manual brake is n...

Page 57: ...s when there is a problem with the Trendelenberg angle sensor Because of this the value for the weight and the angle cannot be displayed When the weight exceeds 550 lbs Figure 50 appears If the weight is less than 2 lbs the screen displays 0 lb When the Trendelenburg or Reverse Trendelenburg angle is above 120 or below 120 Figure 51 appears When the scale system is unstable due to a problem with s...

Page 58: ...wer Sensor board without having any command button pressed The control panel for the weighing scale of the Power Sensor board can also send errors to the micro controller which will display them An error of the weighing scale control panel will light on the Call Maintenance LED ERROR MESSAGES Touch Screen Error Messages Name Definition Atd Invalid Values Digital to analog converter is damaged Repl...

Page 59: ...erload Brake motor is drawing too many AMPS remove restriction or replace motor Motor Foot Overheat Foot motor has ran too long Leave motor stationary for 54 minutes Motor Foot Overload Foot motor is drawing too many AMPS remove restriction or replace motor Motor Gatch Overheat Gatch motor has ran too long Leave motor stationary for 54 minutes Motor Gatch Overload Gatch motor is drawing too many A...

Page 60: ... for switch Verify wiring to Zoom motor Wi Fi IR Communication Error Definition Wireless Adapter Failure Unable to detect or communicate with Wi Fi radio Check assembly or cabling of Wi Fi card in footboard or replace Wi Fi card or footboard DHCP Error Unable to obtain IP Address Bed is not able to connect to the wireless network Check wireless network settings and Wi Fi configuration of bed Unabl...

Page 61: ... Seat Right inflate valve not detected 1104 Seat Right deflate valve not detected 1105 Back Left inflate valve not detected 1106 Back Left deflate valve not detected 1107 Back Right inflate valve not detected 1108 Back Right deflate valve not detected 1109 Turn Left inflate valve not detected 1110 Turn Left deflate valve not detected 1111 Turn Right inflate valve not detected 1112 Turn Right defla...

Page 62: ...ttress leak 1501 Seat Left zone leak 1502 Seat Right zone leak 1503 Back Left zone leak 1504 Back Right zone leak 1505 Turn Left bladder leak 1506 Turn Right bladder leak 1600 Mattress deflate timeout 1601 Seat Left deflate timeout 1602 Seat Right deflate timeout 1603 Back Left deflate timeout 1604 Back Right deflate timeout 1605 Turn Left deflate timeout 1606 Turn Right deflate timeout 1700 Refer...

Page 63: ...Return To Table of Contents www stryker com 2141 209 002 REV H 63 CPU Power Board QDF75 0440 Sanity 5VDC Charge Error ...

Page 64: ... 24 28VDC Pin 1 Black Pin 2 White Bed Lift Foot Down J4 24 28VDC Pin 2 White Pin 1 Black Bed Lift Head Up J4 24 28VDC Pin 1 Black Pin 2 White Bed Lift Head Down J5 24 28VDC Pin 2 White Pin 1 Black Fowler Actuator Up J5 24 28VDC Pin 1 Balck Pin 2 White Fowler Actuator Down J7 24 28VDC 24 28VDC 24 28VDC Pin 1 Black Pin 1 Black Pin 1 Black Pin 2 White Pin 2 White Pin 2 White Brake Brake Off Drive Act...

Page 65: ...OR EXIT 2 1 J1 2 1 J13 EXIT WARNING SIGNAL M M M M M WARNING SIGNAL 2 1 WARNING SIGNAL 6 41 41 6 2 1 FOOT M 2 1 FOOT MOTOR M J3 2 1 1 2 2 1 2 1 1 2 1 2 2 1 1 2 2 1 2 1 J140 J142 J144 J147 J150 5 J131 1 2 2 1 1 2 J149 J131 5 2 1 1 2 6 2 1 1 2 J146 5 J131 3 WARNING SIGNAL FOOT M GATCH M FOOT HI LO MOTOR BRAKE MOTOR ZOOM ACTUATOR HEAD HI LO MOTOR EXIT HEAD MOTOR EXIT FOOT HI LO MOTOR EXIT BRAKE MOTOR...

Page 66: ... R LED_C Y LED_A NC NC J1 J2 J3 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 SCL TPOLT 0V TPOTC TPOTH 5V GND SW24V RAW24V SAFETY_SW J7 M1 2 wires M1 2 wires J5 J10 3 4 5 6 10K THERMOSTAT THERMOSTAT J8 J1 2 1 11 J11 74 7 3 2 3 10 14 16 17 17 16 15 8 8 29 BATTERIES 29 J132 J130 25A Slow blow 25A Slow blow J4 J3 J6 7 19 1 2 J2 J9 1 2 15 5 13 18 16 3 4 SDA J7A J11A J11 J7 11 15 5 13 18 16 3 4...

Page 67: ...und Interchangable plug according to country A B SCHURTER POWER ENTRY MODULE 18 62 20 63 A B MEDICAL FILTER 67 68 J204 Power Supply BLACK WHITE J202 J200 PLACE A LABEL MAX 3A 26 PLACE A LABEL MAX 3A ground J205 73 Mattress Network Plug To J17 73 foot frame ground GATCH FRAME FOOT FRAME HEAD FRAME 56 24 OPTIONAL ground 73 OPTION 120 V DC POWER AND CONTROL CIRCUIT ASSEMBLY TOROID POWER SUPPLY B C AL...

Page 68: ... YELLOW ORANGE 1 2 3 4 Line Neutral Ground Interchangeable plug according to country A B SA PEM SCHURTER 18 62 20 63 A B MEDICAL FILTER 67 68 Power Supply J202 J200 foot frame ground GATCH FRAME FOOT FRAME HEAD FRAME 41 33 J204 BLACK 0 V WHITE 100 V RED 100 V YELLOW 120 V OPTION 100 V DC POWER AND CONTROL CIRCUIT ASSEMBLY MULTIVOLTAGE POWER SUPPLY 44 B C ALL MODELS ...

Page 69: ... 1 2 3 4 A B SA PEM SCHURTER 18 62 20 63 A B 67 68 J204 J202 J200 OPTION 110 V Line Neutral Ground Interchangeable plug according to country HEAD FRAME MEDICAL FILTER Gatch Frame Ground BLACK WHITE BLUE RED YELLOW Foot frame ground FOOT FRAME Power Supply DC POWER AND CONTROL CIRCUIT ASSEMBLY TOROID POWER SUPPLY B C 2131 to 2141 ...

Page 70: ...68 J202 J200 GATCH FRAME HEAD FRAME 41 57 J204 OPTION 200 V Interchangeable plug according to country Line Neutral Ground NEUTRAL LIVE gatch frame ground BLUE 0 V ORANGE 140 V BLACK 0 V WHITE 100 V RED 100 V YELLOW 120 V BLUE RED YELLOW ORANGE Power Supply foot frame ground FOOT FRAME DC POWER AND CONTROL CIRCUIT ASSEMBLY 44 MULTIVOLTAGE POWER SUPPLY B C ALL MODELS ...

Page 71: ... 58 J204 OPTION 220 V Interchangeable plug according to country Line Neutral Ground MEDICAL FILTER HEAD FRAME NEUTRAL LIVE gatch frame ground GATCH FRAME BLUE 0 V ORANGE 140 V BLACK 0 V WHITE 100 V RED 100 V YELLOW 120 V BLUE RED YELLOW ORANGE foot frame ground FOOT FRAME Power Supply DC POWER AND CONTROL CIRCUIT ASSEMBLY MULTIVOLTAGE POWER SUPPLY 44 B C ALL MODELS ...

Page 72: ... J202 J200 26 PLACE A LABEL MAX 3A J205 73 56 OPTION 230V Line Neutral Ground Interchangeable plug according to country HEAD FRAME MEDICAL FILTER Gatch Frame Ground BLACK WHITE BLUE RED YELLOW BLACK WHITE Foot frame ground FOOT FRAME Power Supply GND Mattress Network Plug DC POWER AND CONTROL CIRCUIT ASSEMBLY 230V TRANSFORMER B C Bed Electrical Diagram 2131 to 2156 ...

Page 73: ... 64 J204 OPTION 240 V Interchangeable plug according to country Line Neutral Ground MEDICAL FILTER HEAD FRAME NEUTRAL LIVE gatch frame ground GATCH FRAME BLUE 0 V ORANGE 140 V BLACK 0 V WHITE 100 V RED 100 V YELLOW 120 V foot frame ground FOOT FRAME BLUE RED YELLOW ORANGE Power Supply 44 DC POWER AND CONTROL CIRCUIT ASSEMBLY MULTIVOLTAGE POWER SUPPLY B C ALL MODELS ...

Page 74: ... NC 3 NC LIMIT SWITCH Limit switch 30 J171 LIMIT SWITCH 3 1 NO COM LIMIT SWITCH Limit switch J150 COM NO COM NO 30 3 2 1 25 Analog SIGNAL BRAKE POSITION SIGNAL ZOOM SIGNAL CPR SIGNAL MANUAL ENGAGE 3 2 1 3 2 1 3 2 1 J28 GNDD J116 J170 J127 5 3 2 1 4 3 J10 Cellule de charge N 1 Load Cell N 1 Cellule de charge N 4 Load Cell N 4 Cellule de charge N 3 Load Cell N 3 Cellule de charge N 2 Load Cell N 2 L...

Page 75: ...ake_SW_Common Brake_4SW_BRAKE Brake_SW_Steer Brake_SW_Trend Brake_SW_CPR 12 11 14 13 15 11 15 12 13 14 Brake_LED_Bright 24VA Brake_Led_Yellow Brake_Led_Green Brake_Led_Red J3R1 CAN LO A 12V 24V GNDD CAN HI A SAFE A2 11 TOUCH SCREEN 2 1 3 66 66 J1 J1 73 LEFT SPEAKER RIGHT SPEAKER 8 10 9 7 4 6 5 3 J9 J19 J16 J6 1 2 3 4 5 6 42 7 8 9 13 Brake_SW_Neutral BED CHARGING BRAKE NOT SET BATTERY LOW 24V CALL ...

Page 76: ...ED_Bright Brake_Led_Yellow Brake_Led_Green J3R1 CAN LO A 12V 24V GNDD CAN HI A SAFE A2 11 TOUCH SCREEN 66 J1 J1 73 LEFT SPEAKER RIGHT SPEAKER 8 10 9 7 4 6 5 3 J9 J19 J16 J6 1 2 3 4 5 6 42 7 8 9 13 Brake_SW_Neutral BED CHARGING BRAKE NOT SET BATTERY LOW 24V CALL MAINTENANCE LOCKS ENABLED BTN BLIGHT 5V GND GND 10 11 12 1 2 3 4 5 6 7 8 9 13 10 11 12 BED UNPLUGGED BED EXIT 4 2 1 5 3 USB1_N GND USB1_P ...

Page 77: ..._CPR Brake_LED_Bright 24VA Brake_Led_Yellow Brake_Led_Green Brake_Led_Red 10 11 12 1 2 3 4 5 6 7 8 9 13 14 15 10 11 12 1 2 4 2 1 6 5 3 CAN_L_A 12V 24V GND CAN_H_A SAFE A1 41 4 2 1 6 5 3 CAN_L_A 12V 24V GND CAN_H_A SAFE A1 J20B 11 11 REMOVABLE PENDANT NETWORK J20A LEFT SIDERAIL NETWORK J20C RIGHT SIDERAIL NETWORK 3 4 J5 1 2 3 4 J3 LIMIT SWITCH 1 2 NC NO 3 COM 40 LIMIT SWITCH 3 2 NO NC 1 COM 40 2 1 ...

Page 78: ...t Speaker SPK_HIGH_JACK SPK_LOW_COMMON AUDIO_SHIELD J103 61 41 Left Right J21 1 3 4 2 Left Right Foot 49 Pendant Foot Right NC_NO NCLSR_NC NCLSR_COM 1 2 3 J16 Nurse Call Head Right Siderail 49 41 J102 Right 70 BED STATUS INDICATOR RIGHT siderail 5 6 7 5 6 7 J171 J172 12V GNDD 2 1 3 32 GNDD 2 1 3 12V GNDD 2 1 3 32 GNDD BED STATUS INDICATOR 2 1 3 66 66 41 41 9 9 4 5 6 4 5 6 39 12V 24V GND 12 V NIGHT...

Page 79: ...PEAKER LOW COM TV TV COM TV OPTION 3A COM PRIORITY COM PRIORITY NO NC NURSE ANSWER LIGHT NURSE CALL LIGHT ROOM READ LIGHT COM N_CALL NO NC N_CALL OPTION 2A NO NC OPTION 2A COM OPTION 3A NO NC OPTION 3 NO NC OPTION 2 NO NC NURSE CALL LIGHT OPTION 1 COM OPTION 1 NO NC NURSE ANSWER LIGHT POT HI COM POT LOW COM OPTION 3 COM SPARE INTERLOCK INTERLOCK AUDIO TRANSFERT AUDIO TRANSFERT RADIO COM POT WIPER ...

Page 80: ...O NC N_CALL OPTION 2A NO NC BRAKE STATUS ON OPTION 3A NO NC OPTION 3 NO NC CH DOWN NO NURSE CALL LIGHT IBED SIDERAIL ALERTING IIBED ALERTING NURSE ANSWER LIGHT POT HI COM POT LOW COM OPTION 3 COM IBED FOWLER 30 ALERTING INTERLOCK INTERLOCK AUDIO TRANSFERT AUDIO TRANSFERT BED EXIT STATUS POT WIPER SPEAKER HIGH ROOM LIGHT READ LIGHT IIBED STATUS ON N_CALL INTERLOCK ROOM CONTROL INTERFACE ASSEMBLY IR...

Page 81: ...al Diagram J6 Smart TV 20 10 8 7 9 4 2 1 6 5 3 11 12 13 14 15 16 17 18 19 21 22 23 24 DB_TV DB_TV SMART_TV_ON_OFF SMART_TV_CH_DOWN SMART_TV_CH_UP CLOCK SMART_TV_CC_SLEEP DATA SMART_TV_RADIO_MUTE GND GND GND 39 SMART TV BOARD 52 ROOM CONTROL INTERFACE CIRCUIT ASSEMBLY H ALL MODELS ...

Page 82: ...urse Call Board Cable 2 24 QDF8024 120V Auxiliary Outlet 1 25 QDF27 2024 Linear Position Sensor 1 26 QDF9025 Breaker 1 27 QDF27 2214 Footboard Control Board Cable 1 28 QDF27 2038 Toroid Power Supply 1 29 QDF27 1646 Battery Cable 2 30 QDF9004 Micro Switch 2 31 QDF27 2232 Footboard Cable 1 32 QDF27 1834 Footboard LBS Cable 2 33 QDF27 1841 100V Adapter 1 35 QDF27 1372 Load Cell 4 36 QDF21 1151 CAN He...

Page 83: ...62 Local Bed Status Board 2 67 QDF9571 Medical Filter 1 68 QDF27 1524 Filter and Receptacle Connector 1 69 QDF27 1681 Interior Siderail Cable 2 70 QDF27 1526 Speaker 4 72 QDF27 2231 USB Cable 1 73 QDF27 1976 Auxiliary Plug Mattress Cable 1 74 QDF27 2542 Zoom Interface Board Harness 1 75 QDF75 0240 Zoom Interface Board 1 76 QDF27 2658 230V Transformer 1 77 QDF75 0310 IR Module 2 78 QDF75 0630 Wi Fi...

Page 84: ... all the way down so the litter is supported by the base litter stop 8 If the bed will not lower all the way down using a jack and a 2 x 4 take tension off of the actuator clevis pins by placing the jack and 2 x 4 between the bottom of the right side of the litter and the floor and jack up just enough to take the litter weight off of the base frame reference picture above CAUTION Make sure that pr...

Page 85: ...d out 5 Using diagonal pliers cut the zip tie securing the actuator cable to the base frame 6 Using Needle nose pliers remove the rue clips from the clevis pins securing the actuator CAUTION Do not remove the clevis pins 7 Lower the bed all of the way down so the litter is supported by the base litter stops 8 Using the bed up down controls tap the up or down button to remove tension on the clevis ...

Page 86: ...cut the zip ties securing the CPR cable to the actuator and the litter frame Also cut the zip ties securing the actuator cable to the wire harness 9 Unplug the actuator from the quick connect 10 Holding the actuator with one hand remove the clevis pins from the actuator and carefully lower the actuator down 11 Using a Phillips screwdriver and a 3 8 combination wrench remove the CPR cable retaining...

Page 87: ...ns If the actuator will not run support the gatch section to take tension off of the actuator mounting pins 6 Unplug the actuator cable from the quick connect near the actuator 7 Using Needle nose pliers remove the two rue clips from the clevis pins holding the actuator in place 8 Holding the actuator with one hand remove the clevis pins securing the actuator to the bed then remove the actuator 9 ...

Page 88: ...t section 7 Using a Phillips screwdriver remove the two screws reference drawing above item A and B securing the electrical cover from the foot section and remove the cover Note Use caution as the cover is large and heavy 8 Unplug the actuator from J1 and feed cable down to actuator 9 Using needle nose pliers remove the two rue clips from the clevis pins holding the actuator in place 10 Holding th...

Page 89: ...ng Needle nose pliers remove the two rue clips from the clevis pins securing the actuator to the base frame 5 Using diagonal pliers cut the zip ties securing the actuator cable and then unplug the actuator from the quick connect 6 Using a pry bar or a 2 x 4 pry upward on the Zoom drive actuator frame and push the actuator clevis pins out 7 Remove actuator 8 Reverse steps to install new actuator 9 ...

Page 90: ...stem properly ground yourself 8 Unplug all cable connections from the CPU Power board 9 Using a 3 8 nutdriver and a Phillips screwdriver unscrew the four screws C securing the metal CPU Power board mounting plate to the electrical tray reference drawing above 10 Remove CPU Power board and discard 11 Reverse steps 9 4 to install the new CPU Power board QDF75 0440 12 Plug bed back into wall outlet a...

Page 91: ...nection and feed the connection back towards the cell 7 Using Needle nose pliers remove the rue clip and the flat washer 8 Using one hand grab the upper frame cross bar to remove tension on the litter then push the clevis pin out 9 Using a ratchet with a 1 2 socket and a 1 2 combination wrench remove the two bolts securing the load cell then remove the load cell 10 Reverse procedures to install th...

Page 92: ...e the rue clip and flat washer 7 Remove the foot end base cover 8 Using a Phillips screwdriver remove the four screws securing the foley bag bracket 9 Using a jack and 2 x 4 take the tension off of the load cells and remove the clevis pins from both foot end load cells 10 Unscrew load cell cable connector 11 Using diagonal pliers cut the two cable ties securing the main wire harness and the sidera...

Page 93: ...y remove the display panel from the footboard assembly Rotate over to display the circuit boards 8 Disconnect the two ribbon cables from J3 and J7 by pulling outward on the locking tabs then pulling the ribbon cable 9 Disconnect the menu board cable from connector J6 10 Using a Phillips screwdriver remove the two screws that secure the Wi Fi board to the LCD interface board 11 Using a Phillips scr...

Page 94: ...lay panel up 5 Using a stubby Phillips screwdriver remove the two screws from the bottom of the display panel 6 Hold the display panel and footboard assembly together and flip the footboard assembly over with the display panel facing up 7 Carefully remove the display panel from the footboard assembly Rotate over to display the circuit boards 8 Using a Phillips screwdriver remove the two screws tha...

Page 95: ...ith a new label QDF27 2756 5 Using a Phillips screwdriver remove the six screws from the top of the control panel which the label was covering 6 Raise the display up and using a stubby Phillips screwdriver remove the three screws on the bottom 7 Remove the display housing and unplug the two cables from the function control board and set the display assembly on a bench to work on 8 Using a Phillips...

Page 96: ...by pulling outward on the locking tabs then pulling the ribbon cable 9 Disconnect the menu board cable from connector J6 10 Using a Phillips screwdriver remove the two screws that secure the Wi Fi board to the LCD interface board 11 Using a Phillips screwdriver remove the two screws that secure the LCD interface board to the display housing 12 Remove the two spacer shims between the LCD interface ...

Page 97: ... utility knife remove the rear footboard label Note This label will need to be replaced with a new label QDF27 2756 5 Using a Phillips screwdriver remove the six screws from the top of the control panel which the label was covering 6 Raise the display up and using a stubby Phillips screwdriver remove the three screws on the bottom 7 Remove the display housing and unplug the two cables from the fun...

Page 98: ...d 4 Remove the patient s left base frame cover 5 Cut the two wire ties securing the potentiometer to the potentiometer bracket then cut the two wire ties securing the potentiometer wires to the frame 6 Remove the electrical tape holding the quick connection together 7 Reverse the above procedures for installation of the new potentiometer Note Make sure to install new wire ties in the same location...

Page 99: ...e electrical cover and remove the cover reference Figure 1 above item A and B Note Use caution as the cover is large and heavy 6 Unplug the bed from the wall outlet and turn battery switch OFF 0 The bed should now have no power 7 Unplug J12 connector from the CPU Power board 8 Lift up on both batteries and stand them upright reference Figure 2 above 9 Using a 5 16 nutdriver and a 5 16 combination ...

Page 100: ...motor will not run put the bed into the calibration mode step one as shown on page 25 Run the Fowler up to approximately 20 degrees 5 Working form the left side of the bed use a stubby Phillips screwdriver and remove the two screws securing the fowler angle sensor to the bottom of the fowler frame 6 Using an ESD system properly ground yourself 7 Unclip the three clips holding the board cover on 8 ...

Page 101: ...then put the bed into the calibration mode step one as shown on page 25 5 Working from the bottom left side of the litter under the gatch section use a stubby Phillips screwdriver and remove the two screws securing the gatch angle sensor to the bottom of the gatch frame 6 Using an ESD system properly ground yourself 7 Unclip the three clips holding the board cover on 8 Unplug the cable from the bo...

Page 102: ...remove the cover refer to Figure 1 above item A and B 5 Using a Phillips screwdriver remove the two screws securing the angle sensor to the foot frame refer to Figure 2 above Note Use caution as the cover is large and heavy 6 Using an ESD system properly ground yourself 7 Unclip the three clips holding the board cover on 8 Unplug the cable from the board 9 Reverse the steps to install the new foot...

Page 103: ... end and head end 4 Remove the head end enter foot end and right base cover 5 Using a Phillips screwdriver remove the two screws securing the base angle sensor to the base frame reference Figure above 6 Using an ESD system properly ground yourself 7 Unclip the three clips holding the board cover on 8 Unplug the cable from the board 9 Reverse the steps to install new base angle sensor Note Do not o...

Page 104: ... right side use a Phillips screwdriver and remove the two screws securing the trend angle sensor to the litter frame refer to Figure above 4 Using an ESD system properly ground yourself 5 Unclip the three clips holding the board cover on 6 Unplug the cable from the board 7 Reverse the steps to install new trend angle sensor Note Do not over tighten the two screws 8 Recalibrate the bed refer to Bed...

Page 105: ...ewdriver remove the two screws that secure the right IR module and gasket to the head end frame Note Ensure the IR module cable is routed in the wire track before reinstallation 11 Using a Phillips screwdriver remove the screw that secures the headwall communication board to the head end frame 12 Disconnect all of the cables going to the headwall communication board Note Make note of all cable con...

Page 106: ...ips screwdriver remove the two screws that secure the right IR module and gasket to the head end frame Note Ensure the IR module cable is routed in the wire track before reinstallation 11 Using a Phillips screwdriver remove the two screws that secure the left IR module and gasket to the head end frame 12 Using diagonal pliers cut the cable tie that secures the IR module cables below the head end f...

Page 107: ...sing a small slotted screwdriver or a utility knife remove the rear footboard label Note This label will need to be replaced with a new label QDF27 2756 5 Using a Phillips screwdriver remove the six screws securing the footboard display panel to the footboard 6 Pivot the display panel up and remove the two screws from the bottom of the display panel using a stubby Phillips screwdriver 7 Hold the d...

Page 108: ... s wireless network 17 Configure the bed s Wi Fi settings with hospital network information a Access the Configuration Menu by referencing the procedure on page 25 b On the Configuration Menu screen press Wi Fi Configuration c Press the Advanced tab d Press Reset e Press OK f Press Save g In the pop up message screen press Close h Using the Wi Fi configuration tabs configure the bed to the hospita...

Page 109: ...ble of Contents www stryker com 2141 209 002 REV H 109 Base Assembly Hood L27 026 Rev J Reference Only 10 9 8 6 5 7 A 1 13 3 FOOT HEAD 16 NOTE 1 HAND TIGHTENED THE SCREW HEAD MUST BE WELL SEATED ON THE COVER NOTE 1 ...

Page 110: ...t Stopper Support 2 4 27 2123 Ground Chain 1 5 QDF27 1419 Base Transparent Plate 1 6 QP27 1008 Right Base Cover 1 7 QP27 1009 Left Base Cover 1 8 QP27 1087 Foot Base Cover 1 9 QP27 1093 Center Base Cover 1 10 QP27 1105 Head Base Cover 1 11 QP27 1461 Stopper in Base 4 12 VV23A1G24HL Tapping Screw 4 13 VV83A9G16 Tapping Screw 5 15 VVB4A1O24 Thread Rolling Bolt 4 16 VV84A1I16 Flanged Hex Head Tapping...

Page 111: ...EE DETAIL C SEE DETAIL C See Drawing 27 1660 for details related to the electrical wiring HEAD FOOT 34 34 10 3 22 9 27 23 2 16 18 20 35 40 26 16 18 31 30 27 32 5 28 33 11 36 36 36 36 Use 2 Tyraps to attache the case of the potentiometer to the steel support 27 37 GREASE 38 39 DETAIL C To assemble the casters direct the smile chrome part towards the line of the bed ...

Page 112: ... 002 REV H www stryker com Base Assembly Brake DETAIL A 1 27 16 24 8 18 13 14 29 18 35 12 19 36 View of Assembled Brake System DETAIL B The angle sensor orientation is very important Wire should exit towards the bottom 4 27 Tyrap positioning ...

Page 113: ...452Z Microswitch Bracket 1 15 VW10C202802 Washer 4 16 VG50A1224 Clevis Pin 6 17 QDF27 1227 Brake Actuator 1 18 QDF7878 Fastener Pin 10 19 QDF9159 Cherry Limit Switch 1 20 27 2519P Emergency Pedal 1 21 QPA27 1335 Brake Lever 4 22 QRE27 1844 Extension Spring 1 23 VG50A1259 Clevis Pin 1 24 VG50B1236 Clevis Pin 1 25 VG50B1248 Clevis Pin 1 26 VV10A0G16 S Hexagon Cylinder Head 4 27 VV83A9G16 Tapping Scr...

Page 114: ...06 Rev D Reference Only BRAKING NEUTRAL HEAD FOOT BRAKING DIRECT WHEEL 1 1 2 DETAIL A 2 DETAIL A LEFT RIGHT DETAIL A To assemble casters direct the smile chrome part toward the left side of the bed Item Part No Part Name Qty 1 RD27 2787 5 5 Caster Brake Neutral Neutral 2 2 RD27 2788 5 5 Caster Brake Neutral Drive 2 ...

Page 115: ...KING NEUTRAL SEE DETAIL A SEE DETAIL A 1 1 1 1 LEFT RIGHT Base Assembly Caster Non Lock Model 2141 Only OL270005 Rev D Reference Only DETAIL A Item Part No Part Name Qty 1 RD27 2787 5 5 Caster Brake Neutral Neutral 4 To assemble casters direct the smile chromed part toward the left side of the bed ...

Page 116: ... 116 2141 209 002 REV H www stryker com Angle Sensor Assembly ISO 1 2 3 Item Part No Part Name Qty 1 QDF75 0140 Angle Sensor 1 2 QP27 1210 Angle Sensor Case 1 3 QP27 1211 Bottom Lid for Tilt Sensor 1 27 2477 Rev 0 Reference Only ...

Page 117: ... 2141 209 002 REV H 117 Base Assembly Electrical 27 2687 Rev B Reference Only 2 J131 2 J127 2 J147 2 J146 2 J149 2 J170B 2 J170A 3 3 3 3 HEAD FOOT 2 J116 2 J115 3 3 see photo 9 see photo 4 2 J126 see photo 2 and 3 see photo 10 see photo 1 and 14 ...

Page 118: ... 002 REV H www stryker com Base Assembly Electrical 1 J126 1 J131 1 J115 3 3 3 J4 J6 J7 J30 J24 toward DC Control Board see drawing Litters Wiring for opening in litter 3 Leave 2 Loop t t A B see photo 13 see photo 11 see photo 6 see photo 1 ...

Page 119: ... REV H 119 Base Assembly Electrical 3 3 3 Leave 2 loop 3 A B see photo 8 see photo 5 see photo 7 and 12 Note To avoid any risk of pinching when assembling these cables Bed must be in high position Gatch must be at 15 degrees Foot Litter must be at 45 degrees ...

Page 120: ...ents 120 2141 209 002 REV H www stryker com Base Assembly Electrical PHOTO 1 Be sure to place the connectors one beside the other and in a flat position PHOTO 2 PHOTO 3 PHOTO 5 PHOTO 6 PHOTO 7 PHOTO 8 PHOTO 9 PHOTO 10 PHOTO 4 ...

Page 121: ...81 J126 Hi lo Harness QDF27 2182 J131 To 27 1181 J131 Hi lo Harness QDF27 2182 J115 To 27 2277 Sensor Base QDF27 2182 J127 To QDF2083 CPR Switch QDF27 2182 J116 To QDF27 2724 Brake Potentiometer QDF27 2182 Bornes J170A To QDF9159 No Manual Switch QDF27 2182 Bornes J170B To QDF9159 COM Manual Switch QDF27 2182 J146 To QDF27 1252 Hi lo Head Motor QDF27 2182 J147 To QDF27 1251 Hi lo Foot Motor QDF27 ...

Page 122: ...y Electrical CABLE QDF27 2181 CONNECTION TABLE Cable No Connector No To Cable No Connector No QDF27 2181 J126 To QDF27 2182 J126 Hi lo Harness QDF27 2181 J131 To QDF27 2182 J131 Hi lo Harness QDF27 2181 J115 To 27 2477 Sensor Base QDF27 2181 J7 To QDF75 0440 J7 DC Control Board QDF27 2181 J6 To QDF75 0440 J6 DC Control Board QDF27 2181 J4 To QDF75 0440 J4 DC Control Board QDF27 2181 J30 To QDF75 0...

Page 123: ...onnector No QDF27 1646 Mini Fit To QDF27 1185 J132 QDF27 1646 Black Eyelets To QDF9188 BAT 2 POLE QDF27 1646 White Eyelets To QDF9188 BAT 1 POLE 1 J27 J8 TO PC BOARD DC DANS 27 2478 5 J1 J2 J7 J8 J9 J11 5 J150 5 P3 4 5 J132 3 J132 3 TO BATTERY 2 4 FOR OTHER CABLE CONNECTIONS REFER TO THE ELECTRICAL BASE ASSEMBLY 27 2687 NOTE OL270232 Rev C Reference Only ...

Page 124: ...sembly Zoom QDF27 1185 CONNECTIONS Cable No Connector No To Cable No Connector No QDF27 1185 J27 To QDF21 1174 DC Card J27 QDF27 1185 J8 To QDF21 1174 DC Card J8 QDF27 1185 J132 To QDF27 1646 Mini Fit QDF27 1185 P3 To QDF27 1430 P3 27 2547 XXX QDF27 1185 J150 To QDF27 1445 Zoom 27 2546 QDF27 1185 J1 To QDF27 2548 J1 27 2546 QDF27 1185 J9 To QDF27 2548 J9 27 2546 QDF27 1185 J2 To QDF27 2548 J2 27 2...

Page 125: ... Zoom Drive Model 2141 Only 27 2546 Rev 02 Reference Only 17 17 4 24 7 25 1 8 3 14 2 26 18 17 14 20 19 15 17 9 6 11 13 24 23 16 12 22 21 5 10 J3 J4 J5 J6 J10 28 27 J150 PLACE THE HYDRAULIC GREASER TOWARD THE BOTTOM AND REMOVE THE HEX HEAD SCREW ON THE BEARING FOOT HEAD 29 30 J10 30 ...

Page 126: ...B1248 Clevis Pin Diameter 1 22 VV10A1P20 S Cylindric Head Screw 3 23 VV83A9G16 Pan Head Tapping Screw 2 24 VVB4A1024 Thread Rolling Bolt 6 25 VW10B264004 Steel Washer 1 26 VW10C122802 Nylon Washer 1 27 QDF2134 Board Support 4 28 QDF75 0240 Zoom Interface Board 1 29 QDF27 2542 Zoom Interface Board Harness 1 30 QDF8047 Jumper 1 Base Assembly Zoom Drive Model 2141 Only ACTUATOR CONNECTION QDF27 1445 ...

Page 127: ...Return To Table of Contents www stryker com 2141 209 002 REV H 127 Bed Labeling OL270016 XXX Rev P Reference Only 8 9 12 2 10 3 1 5 6 14 RIGHT SIDE OF THE BED 16 4 BELOW BED 15 ...

Page 128: ...Return To Table of Contents 128 2141 209 002 REV H www stryker com Bed Labeling 19 90 23 95 13 35 16 75 6 5 ABOVE BED 9 12 7 5 8 3 10 13 LEFT SIDE OF THE BED HEAD SIDE 18 ...

Page 129: ...Table of Contents www stryker com 2141 209 002 REV H 129 Bed Labeling 4 8 66 6 38 15 0 83 4 52 22 25 2X 46 07 2X 0 81 2X 1 33 2X 3 10 2 65 0 50 0 39 9 24 7 5 3 93 3 31 13 16 16 26 13 34 SEE DETAIL A 0 90 0 31 14 ...

Page 130: ...tomer Manual Label 2 6 QE71 1207 XXX Electric Shock Hazard Label 1 7 QE71 0943 XXX Grounding Reliability Label 1 8 QE71 0944 XXX Oxygen Tent Label 2 9 QE71 0949 XXX Maximum Mattress Thickness Label 2 10 QE71 0963 XXX CPR Label 2 12 QE71 1010 550 lbs Safe Working Load label 2 13 QE71 0571 Fuse 10A 250V Label 1 14 QE71 1094 XXX Patent Pending Label 1 15 QE71 1042 XXX On Off Label 1 16 QE71 1257 XXX ...

Page 131: ...ntents www stryker com 2141 209 002 REV H 131 Lift Assembly L27 044 Rev C Reference Only 29 FOOT HEAD B 34 47 49 13 48 47 12 11 51 51 52 DETAIL C C 49 13 52 NYLON WASHER TOWARD EXTERIOR OF BED 32 52 DETAIL B 40 19 40 35 42 ...

Page 132: ...H www stryker com Lift Assembly TO ASSEMBLE THIS PART THE ARROW DIRECTED TOWARDS THE HEAD OF THE BED 2 27 RIGHT LEFT 25 10 31 5 7 38 39 33 20 46 45 45 44 28 43 5 41 41 41 INTERIOR OF THE SHOULDERS FOOT HEAD 36 16 18 36 15 19 14 LEFT RIGHT 17 Head End ...

Page 133: ... 39 33 9 4 30 31 26 27 A 32 6 5 1 46 45 44 45 43 41 TO ENSURE ITSELF TO COAT THE SPRING OF GREASE COMPLETELY 41 INTERIOR OF THE SHOULDERS 41 41 37 HEAD FOOT 22 24 18 14 21 23 LEFT RIGHT 19 15 36 Foot End DETAIL A THE ORIENTATION OF THIS SENSOR OF ANGLE IS IMPORTANT THE WIRE MUST LEAVE DOWNWARDS 50 8 37 ...

Page 134: ...4 21 QP27 1236 10 Exterior Right Foot High Lift Cover 1 22 QP27 1237 10 Interior Right Foot High Lift Cover 1 23 QP27 1238 10 Exterior Left Foot High Lift Cover 1 24 QP27 1239 10 Interior Left Foot High Lift Cover 1 25 QPA27 1024 Right Lever Support 2 26 QPA27 1030 Right Stiffener Arm 2 27 QPA27 1036 Left Lever Support 2 28 QPA27 1040 Left Stiffener Arm 2 29 VB18A1O32 Bolt 12 30 VB35A1O40 Carriage...

Page 135: ...07P Spring Support Assembly 2 6 27 2115 Hi Lo Spring Support Glider 2 7 27 2116 Acetal Washer 2 8 M0019 Grease 1 9 QDF7878 Clevis Pin Diameter 6 10 QRC27 2114 Base Compression Spring 2 11 VB15A1O24 Bolt 8 12 VE30A1O Nylon Locknut 8 13 VG50A1228 Clevis Pin 2 14 VG50A1244 Clevis Pin 2 15 VG50A1240 Clevis Pin 2 16 VV83A9G12 Phillips Head Tapping Screw 4 17 27 2139P Base Wiring Routing 2 1 2 3 6 14 9 ...

Page 136: ...81 CONNECTOR J112 27 2256 34 29 36 42 8 B 28 15 36 26 41 36 20 32 D DETAIL D 49 48 64 36 36 45 8 51 10 1 45 C DETAIL C INSTALL PART WITH HOLE IN LOWER POSITION 55 22 42 57 60 37 27 T220 20 THE ORIENTATION OF THIS ANGLE SENSOR IS IMPORTANT THE WIRE MUST EXIT DOWNWARDS 54 6 DETAIL B 37 59 27 39 28 26 50 40 52 61 56 58 58 62 L27 056 Rev E Reference Only ...

Page 137: ... 19 38 41 40 44 43 12 31 TO QDF75 0110 J1 27 2478 TO QDF27 2213 J204 27 2256 TO QDF27 2213 J144 27 2256 TO QDF75 0110 J205 27 2478 TO QDF27 1996 J11 27 2256 46 26 11 63 47 3 SEE OL270210 OL270211 OL270212 OL270213 OL270214 42 35 36 DETAIL A THE ORIENTATION OF THIS ANGLE SENSOR IS IMPORTANT THE WIRE MUST EXIT DOWNWARDS 6 42 ...

Page 138: ...ntermediary Section Cover 1 22 27 2141 Unbalanced Spring Support Rod 2 23 QDF27 1214 Fowler Actuator 1 24 QDF27 1215 Gatch Actuator 1 25 QDF27 1216 Foot Actuator 1 26 QDF7878 Cotter Pin 9 27 QDF7898 Cotter Pin 2 28 QDF7899 Cotter Pin 3 29 QDF9518 Cable Tie 1 31 QP27 1958 Gatch Pivot Stopper 1 32 QRT27 2020 Torsion Spring 1 33 QRT27 2026 Right Upper Frame Spring Base Rod 1 34 VB35A1O32 Carriage Bol...

Page 139: ...nut 2 53 VB35A1O36 Carriage Bolt 2 54 VV33A0G24 Phillips Screw 2 55 VVB4A1O32 Hex Head Thread Rolling Bolt 2 56 27 2223Z Finger 1 57 27 2217 PVC Slide 1 58 27 2227 Bushing Latch Finger 2 59 QRE27 2218 Finger Bracket Spring 1 60 27 2219Z Plate of Tightening 1 61 27 2221 Female Bushing Latch Finger 1 62 27 2220 Male Bushing Latch Finger 1 63 QDF2132 Clevis Pin 1 64 VVB4A1O28 Thread Rolling Bolt 2 ...

Page 140: ... Litter Assembly Interior Frame L27 055 Rev D Reference Only A DETAIL A FOOT HEAD 6 14 19 19 14 20 3 18 12 2 18 12 18 9 11 20 4 5 PASS THE WIRE OVER THE RUBBER SLEEVE 20 5 16 15 13 12 8 17 31 B B SECTION B B 28 GREASE 5 23 24 25 26 27 24 27 25 23 4 9 9 14 29 29 30 30 ...

Page 141: ...Return To Table of Contents www stryker com 2141 209 002 REV H 141 Litter Assembly Interior Frame 10 21 6 15 1 7 10 FOOT HEAD ...

Page 142: ...2 S Bolt 2 11 VB15A1O48 Bolt 2 12 VB15A1O54 Bolt 6 13 VB15A1O56 Hex Bolt 4 14 VE30A1O Nylon Hex Locknut 10 15 VV83A9G16 Pan Head Tapping Screw 6 16 27 1806P Right Frame Support Cover 1 17 27 1385P Left Frame Support Cover 1 18 VW10A10 Flat Washer 10 19 VW20A10 Spring Washer 2 20 VE78A1O Flanged Locknut 8 21 VV83A9G24 Pan Head Tapping Screw 4 23 27 2143P Spacer Foot 2 24 27 2144P Spacer Head 2 25 V...

Page 143: ... Port OL270278 Rev A Reference Only Litter Assembly with Serial Port OL270278 Rev A Reference Only Item Part No Part Name Qty 1 27 1520P Support Connector 1 2 QDF2056 Standoff 2 3 QDF27 2432 Serial Interconnection Cable 1 4 VV83A9G16 Pan Head Phillips Screw 2 CONNECT ON J1 TO QDF75 0270 3 1 2 4 ...

Page 144: ...Return To Table of Contents 144 2141 209 002 REV H www stryker com Litter Assembly Upper Frame L27 046 Rev 01 Reference Only A HEAD FOOT 14 1 5 12 10 15 8 3 3 14 4 7 6 13 13 9 11 16 14 ...

Page 145: ...ire Cover 1 7 27 1851P Right Head Frame Wire Cover 1 8 QDF7878 Coupling Pin 4 9 QE71 1181 Serial Number Plate 1 10 VG50B1248 Clevis Pin 4 11 VR11H43 Pop Rivet 2 12 VV83A9E12 Pan Head Tapping Screw 6 13 VV83A9G12 Pan Head Tapping Screw 4 14 VV83A9G16 Pan Head Tapping Screw 16 15 VW10A12 Washer 4 16 27 2005P Hiding place cell 2 17 VV33A0G24 Phillips Head Screw 2 18 VE30A0G Nylon Nut 2 Litter Assembl...

Page 146: ...ents 146 2141 209 002 REV H www stryker com Litter Assembly Foot End 27 2688 Rev B Reference Only SEE ASSEMBLY OF THE COVER FOOT HEAD 6 33 22 29 10 25 24 30 18 4 22 16 18 16 7 5 21 12 15 14 11 15 21 2 COVER ASSY 4 00 1 00 33 ...

Page 147: ...w stryker com 2141 209 002 REV H 147 Litter Assembly Foot End B 9 23 1 35 17 17 19 20 9 DIRECT WIRE TOWARDS THE INTERIOR OF THE BED 17 31 32 DETAIL B THE ORIENTATION OF THIS ANGLE SENSOR IS IMPORTANT THE WIRE MUST BE DOWNWARDS 3 22 ...

Page 148: ...Return To Table of Contents 148 2141 209 002 REV H www stryker com Litter Assembly Foot End 8 17 22 13 20 26 22 27 ALWAYS ASSEMBLE THE TWO SPACES W O LEG TOWARDS THE INTERIOR OF THE BED 28 34 ...

Page 149: ...x 1 14 QPA27 1576 W Molded Foot Litter End 1 15 VB15A1O32 S Scotch Grip Bolt 2 16 VB15A1O54 Full Thread Bolt 4 17 VE30A0G Nylon Locknut 9 18 VE30A1O Nylon Locknut 4 19 VE80A0G Locknut 1 20 VV33A0G28 Phillips Machine Screw 5 21 VV83A9E16 Phillips Screw 4 22 VV83A9G16 Phillips Screw 11 23 VV87A9A24 Phillips Tapping Screw 4 24 VVB4A1024 Thread Rolling Bolt 4 25 VV83A9G24 Phillips Tapping Screw 4 26 Q...

Page 150: ... Table of Contents 150 2141 209 002 REV H www stryker com Litter Assembly Standard Cabling L27 059 Rev B Reference Only 2 1 Item Part No Part Name Qty 1 27 2687 Base and Lever Wiring 1 2 27 2695 Litter Wiring 1 ...

Page 151: ...12 J11 J10 J9 HEAD CONNECTIONS 7 7B 7A 2 2 7B 7A 11 FILTER LOAD LINE L G N N L BLACK WHITE WHITE BLACK L N G SEE PICTURE 1 AND 6 SEE PICTURE 1 AND 6 12 13 13 11 14 MAKE A LOOP THERE AND TIED WITH A TYRAP SEE PICTURE 3 GREEN OR YELLOW GREEN HEAD LEFT SIDERAIL LEFT FOOT SIDERAIL LEFT FOOT CELL PENDANT RIGHT FOOT SIDERAIL RIGHT FOOT CELL HEAD RIGHT SIDERAIL HEAD SIDE CELL ROOM Note To avoid any bindi...

Page 152: ...ER 4 CARD J12 10 CARD J20 10 LIGHT 4 BAT 4 BAT 9 TRANSFO J205 9 GROUND J202 9 POWER ET DISJ 9 9 DISJ BLACK J206 9 DISJ J207 9 POWER GREEN J208AG 9 POWER J208AL 9 POWER WHITE J208AN HARNESS 1 STARTING POINT 8 CARD J18 8 KNEE GATCH AND FOOT GROUNDING CONNECTIONS GREEN OR GREEN YELLOW WIRE 2 J113 AUXILIARY POWER B A 11 SEE PICTURE 2 12 13 13 15 SEE PICTURE 5 J208A BL J208A BN J208A BG J206 J207 CONTA...

Page 153: ...mmet To QDF9188 Battery QDF27 1381 Terminal To QDF2089 Interrupter QDF27 1381 Terminal To QDF2089 Interrupter QDF27 1646 CABLES 4 CONNECTION TABLE Cable No Connector No To Cable No Connector No QDF27 1646 Mini Fit To QDF75 0440 J12 DC Board QDF27 1646 Black Grommet To QDF9188 Battery 2 Pole QDF27 1646 White Grommet To QDF9188 Battery 1 Pole Photo 1 Photo 2 Photo 3 Photo 4 Photo 5 19 Photo 5 ...

Page 154: ...round QDF27 2213 J114 To QDF75 0140 Trend Sensor QDF27 2213 J153 To QDF27 1372 Right Foot Cell QDF27 2213 J103 To 27 2598 Removable Control QDF27 2213 J172 To QDF27 1208 Right Foot Siderail QDF27 2213 J102 To QDF27 2212 Right Head Siderail QDF27 2213 J152 To QDF27 1372 Right Head Cell QDF27 2213 J202A To Foot Ground QDF27 2213 J16 To QDF75 0270 J16 QDF27 2213 J15 To QDF75 0270 J15 QDF27 2213 J14 T...

Page 155: ...Board QDF5095 Buzzer To Electrical Box See Position on Drawing QDF27 1524 CABLES 7 CONNECTION TABLE Cable No Connector No To Cable No Connector No QDF27 1524 Black 90º Terminal To QDF9571 Line L QDF27 1524 White 90º Terminal To QDF9571 Line L QDF27 1524 Green 90º Terminal To QDF9571 Line L QDF27 1524 Black Right Terminal To 27 2848 L Power Supply QDF27 1524 White Right Terminal To 27 2848 N Power ...

Page 156: ... Assembly Electrical Litter Assembly Electrical 27 2695 Rev B Reference Only Item Part No Part Name Qty 1 QDF27 1381 Battery Switch Wire 2 2 QDF27 2213 1 Wire Harness 1 3 QDF27 1607 I V Pole and Bed Extender Cable 1 4 QDF27 1646 Battery Wires 1 5 QDF5095 Buzzer 1 6 QDF9518 Cable Tie 51 7 QDF27 1524 Filter and Receiver Connector 1 8 QDF27 2214 Panel Control Board Cable 1 9 QDF27 2228 Auxiliary 120V...

Page 157: ...sh French Item Part No Part Name Qty 1 QDF27 2038 Medical Toroidal Transformer 1 2 QE71 1183 XXX 120V CSA Label 2131 Models Only 1 or QE71 1188 XXX 120V CSA Label 2141 Models Only 1 2 OL270206 XXX Rev E Model 2131 Only For Reference Only OL270211 XXX Rev E Model 2141 Only 1 2 1 To Harness 1 27 2213 J204 To DC Control Board J11 To Auxiliary Outlet 27 1976 J205 ...

Page 158: ...Return To Table of Contents 158 2141 209 002 REV H www stryker com Coiled Power Cord 27 2782 Rev A Reference Only ...

Page 159: ... Screw 2 5 QDF9025 5A Breaker 1 6 QDF9573 Single Leviton 8310 G Outlet 1 7 27 1967P Auxiliary Outlet Box 1 Litter Assembly Foot End with Mattress Connector MATTRESS CONNECTOR ASSEMBLY MATTRESS CONNECTOR GROOVE TOWARD INSIDE BOTTOM VIEW QDF27 2248 CABLES 1 CONNECTION TABLE Cable No Connector No To Cable No Connector No QDF27 2248 J17 To QDF27 1960 J17 DC Board QDF27 2248 J17A To Hole 27 1200 OL2701...

Page 160: ...8 10A 250V Fuse 2 2 QDF9571 Medical Filer 1 3 27 2848 Inlet Filter Assembly 1 4 QDF9575 Fuse Holder 1 5 VE30A0G Nylon Hex Locknut 2 6 VV41A1A20 Tapping Screw Flat Head Phillips 2 7 VV83A9G16 Tapping Screw Pan Head Phillips 2 8 27 1723P Head Plate without Auxiliary Outlet 1 9 QDF2155 Steel Cable Clamp 1 10 VV33A0G16 Machine Screw Pan Head Phillips 2 11 VV83A9G24 Pan Head Tapping Screw 1 ...

Page 161: ... TO BROWN GROUND DETAIL A QDF27 2228 QDF27 2228 CABLE CONNECTION TABLE Cable No Connector No Cable No Connector No QDF27 2228 J108A L Black QDF9573 L QDF27 2228 J108A N White QDF9573 N QDF27 2228 J108A G Green QDF9573 GROUND QDF27 2228 J108B L Orange QDF8024 L QDF27 2228 J108B N Yellow QDF8024 N QDF27 2228 J108B G Brown QDF8024 GROUND OL270013 XXX Rev 04 Reference Only ...

Page 162: ... language choice tri English French Spanish bil English French Item Part No Part Name Qty 1 27 1978P Optional Auxiliary Outlet Box 1 2 QDF8024 Leviton 8200 w Double Outlet 1 3 VV37A1C12 Truss Head Machine Screw 1 4 VV83A9G24 Tapping Screw 3 5 QDF2124 Terminal 10 for 16 14 AWG 1 6 QE71 0964 XXX Auxiliary Plug Label XXX 1 Litter Assembly Optional Dual 120V Outlet ...

Page 163: ...Return To Table of Contents www stryker com 2141 209 002 REV H 163 CPR Assembly Mechanical 18 9 19 5 12 4 1 10 13 2 15 17 11 14 16 7 6 5 12 4 3 19 HEAD FOOT L27 033 Rev 09 Reference Only ...

Page 164: ...8 8 8 8 8 8 See Photo 6 See Photo 5 See Photo 4 See Photo 1 2 3 See Photo 7 and 8 DETAIL A 8 8 Leave a 4 loop here 1 00 0 06 Leave a 4 loop here NOTE For Cable Routing The Fowler must at 30 when installing the cable There must be no tie installed on the actuator See Photo 11 See Photo 10 See Photo 9 ...

Page 165: ...Return To Table of Contents www stryker com 2141 209 002 REV H 165 CPR Assembly Mechanical Photo 1 Photo 2 Photo 3 Photo 4 Photo 5 Photo 6 ...

Page 166: ...Return To Table of Contents 166 2141 209 002 REV H www stryker com CPR Assembly Mechanical Photo 7 Photo 8 Photo 9 Photo 10 Photo 11 ...

Page 167: ... 27 2086Z CPR Switch Support 1 7 QDF2083 Lever Switch 1 8 QDF9518 Cable Tie 12 9 QDF2155 Steel Cable Clamp 3 10 QP19 0270 Shaft Support Bearing 4 11 QP27 2065 CPR Pedal 2 12 QRE23 0438 Siderail Spring 2 13 VB15A1N44 Bolt 2 14 VE30A0G Nylon Locknut 1 15 VE30A1N Nylon Locknut 2 16 VG10B0636 Spring Pin 2 17 VV33A0G24 Pan Head Machine Screw 1 18 VV83A9G12 Pan Head Tapping Screw 3 19 VV83A9G16 Pan Head...

Page 168: ...11 11 Item Part No Part Name Qty 1 27 1651P Without 5th Wheel Plate 1 2 VV83A9G12 Pan Head Tapping Screw 3 3 QDF5096 Plastic Tie 2 5 VE30A1O Nylon Hex Locknut 4 6 VB15A1O32 Bolt 4 7 27 2790P Zoom Handle Bracket Right 1 8 27 2789P Zoom Handle Bracket Left 1 11 QE71 1373 Coiled Power Cord Label 2 OL270058 Rev G Reference Only ...

Page 169: ...ontents www stryker com 2141 209 002 REV H 169 Zoom Handle Assembly Model 2141 Only A A 13 6 8 3 7 2 1 9 4 14 15 18 17 16 22 19 12 20 10 10 20 26 23 26 27 2547 XXX Rev B Reference Only Note Attach cables under the head box ...

Page 170: ...80C Tie Bus 1 4 27 1181 Zoom Handle Sleeve 3 6 27 2548 Zoom Control Board Assembly 1 7 QDF9136 250 lbs Load Cell 1 8 VB15A1O32 Bolt 4 9 VB95A1P48 Hex Neck Bolt 2 10 VE42A9036 Combination Barrel Nut 2 12 VE30A1O Nylon Locknut 4 13 VE30A1P Nylon Locknut 2 14 VV83A9G12 Phillips Head Tapping Screw 3 15 VV10A0G16 S Hex Head Screw 2 16 27 1857Z Bottom Zoom Load Cell Stopper 1 17 QDF27 1780 Load Cell Spa...

Page 171: ...209 002 REV H 171 1 2 TOP VIEW 2 1 Zoom Assembly NOTE XXX indicates language choice tri English French Spanish bil English French Item Part No Part Name Qty 1 QE71 0946 XXX Zoom Label 1 2 QE71 1372 Zoom Handle Label 1 OL270077 XXX Rev D Reference Only ...

Page 172: ...ead End Control Panel 1 3 2 No Head End Control Panel Assembly OL270019 Rev 01 Reference Only Item Part No Part Name Qty 1 QDF27 1649 Head End Control Fascia without Motorized Zoom Drive 1 2 QP27 1650 Head Frame Plate 1 3 VV23A9C12HL Pan Head Screw 4 OL270019 Rev 01 Reference Only ...

Page 173: ...n To Table of Contents www stryker com 2141 209 002 REV H 173 Optional Head End Control Panel Assembly 2 5 6 TOWARDS J1 QDF75 0410 1 J6 4 3 7 Brake Off Drive N QDF27 2236 3 OL270263 XXX Rev C Reference Only ...

Page 174: ... 2 QDF27 1099 Siderail Nurse Control Board 1 3 QF27 2236 XXX Fascia Head of Bed Control 1 4 QDF9183 Electronic Board Button 12 5 QP27 1111 Nurse Control Board Support 1 6 VV23A9C12HL Pan Head Screw 10 7 QE71 1340 Connection Isolation Sticker 1 QDF21 2895 CONNECTION TABLE Cable No Connector No Cable No Connector No QDF21 2895 MTA6 QDF27 1099 J6 QDF21 2895 MTA6 QDF75 0270 J1 DIP SWITCH CONFIGURATION...

Page 175: ...41 209 002 REV H 175 Item Part No Part Name Qty 1 VB18A1O32 Bolt 8 2 VE30A1O Nylon Locknut 8 3 27 2287 Head End Siderail Right 1 4 27 2288 Head End Siderail Left 1 Siderail Mounting Assembly Head End L27 039 Rev 01 Reference Only 4 3 1 2 Head Foot ...

Page 176: ...nts 176 2141 209 002 REV H www stryker com 44 20 21 26 8 6 23 19 1 30 7 14 B A 39 46 LUBRICATE 34 25 31 29 T35 3 5 T22 2 Standard Siderail Assembly Head End Right 27 2287 Rev 04 Reference Only DETAIL A 28 5 16 39 B 27 40 45 4 ...

Page 177: ...H 177 Standard Siderail Assembly Head End Right TOWARDS LITTER 35 24 SEE NOTE 1 32 36 39 J5 35 J6 38 J3 38 J1 38 10 31 17 18 3 20 B 31 31 31 9 35 J2 ET MTA 3 41 42 43 15 SEE SIDERAIL OPTION 47 PART QP28 0040 IS OF GRAY COLOR CODE OF COLOR 10 NOTE 1 QDF27 2237 10 ...

Page 178: ...Green 3 3 N C Brown 1 1 Speaker Blue 2 2 Speaker Orange DIP SWITCH SETUP ON QDF27 1099 SW13 ON ON ON SW14 ON ON CABLES CONNECTION TABLE Cable No Connector No To Cable No Connector No QDF27 1681 MTA 3 Positions To QDF27 1099 J5 QDF27 1156 MTA 6 Positions To QDF27 1099 J3 QDF27 1156 MTA 6 Positions To QDF27 1097 J1 QDF27 1784 MTA 3 Positions To QDF27 2212 MTA 2 Positions QDF27 2212 Mate N Lock 12 Po...

Page 179: ...l Board Support 1 19 QP27 2225 05 Right Head Siderail 1 20 QP27 1265 05 Right Head Siderail Cover 1 21 QP27 1273 08 Right Siderail Handle 1 23 QP27 1496 05 Head Siderail Cap 1 24 QP28 0040 10 Reversed Siderail Arm Cover 2 25 QPA28 0015 Siderail Sleeve 1 26 QPA28 0016 Sleeve with Lock 1 27 QRE27 1736 Return Spring 1 28 VV10A0G16 S Cylinder Head Screw 2 29 VV10B0G40 S Cylinder Head Screw 4 30 VV10A1...

Page 180: ...ontents 180 2141 209 002 REV H www stryker com Standard Siderail Assembly Head End Left 27 2288 Rev 04 Reference Only 22 18 7 6 20 25 12 1 27 A 39 26 4 5 B 46 LUBRICATE 30 29 T35 3 5 28 7 34 24 T22 2 DETAIL A 15 39 40 B 45 ...

Page 181: ...EV H 181 Standard Siderail Assembly Head End Left PART QP28 0039 IS OF GRAY COLOR CODE OF COLOR 10 NOTE 1 TOWARDS LITTER 36 35 J1 35 J3 36 J6 39 J5 9 14 17 23 SEE NOTE 1 32 37 B 3 36 MTA 2 31 14 16 8 36 J2 19 41 43 42 31 SEE SIDERAIL OPTION 44 47 QDF27 2238 9 ...

Page 182: ... 038 QDF27 2212 MTA 6 Positions To QDF27 1099 J6 QDF27 2212 MTA 3 Positions To See OL270162 OL270163 OL270164 OL270173 OL270174 OL270175 QDF27 2212 MTA 2 Positions To QDF27 2212 Connector 2 Pins To POSITION ON CONNECTOR Cable 36 Position on Connector 47 Position on Connector 41 Position on Connector 43 Position on Connector 42 Signal Color 1 1 CAN H Black Red 2 2 CAN L Red 3 3 12V Black White 4 4 ...

Page 183: ... Siderail Cover 1 20 QP27 1274 08 Left Siderail Handle 1 22 QP27 1496 05 Head Siderail Cap 1 23 QP28 0039 10 Siderail Rotation Lever Cover 2 24 QPA27 2296 Left Head Siderail Structure 1 25 QPA28 0015 Siderail Sleeve 1 26 QPA28 0016 Sleeve with Lock 1 27 QRE27 1736 Spring Balance 1 28 VV10A0G16 S Hexagon Socket Head Screw 2 29 VV10B0G40 S Hexagon Socket Head Screw 4 30 VV10A1P20 S Hexagon Socket He...

Page 184: ...XX Nurse Control Label HE Right 1 2 QDF27 0907 XXX Nurse Control Label HE Left 1 3 QDF27 2239 XXX Brake Control Label HE Right 1 4 QDF27 2240 XXX Brake Control Label HE Left 1 Brake Set Brake Not Set QDF27 2239 Vascular Position Brake Off Drive N QDF27 0906 WhiteInteriorwill Glowinthedark 3 1 4 2 Head End Left Head End Right 1 2 Right label shown QDF27 0906 3 4 Right label shown QDF27 2239 OL27018...

Page 185: ...r 2 7 QDF27 2216 Siderail Speaker Right 1 8 QDF27 2210 Label Right Communications 1 27 2212 5 6 2 3 3 27 2212 Cable No Cable Connection To Cable No Cable Connection QDF27 2175 MTA 2 Positions To QDF27 2212 MTA 2 Positions Standard Siderail Assembly Head End Left QDF27 2216 MTA 2 Positions To QDF27 2212 MTA 2 Positions Standard Siderail Assembly Head End Right QDF27 2196 2 Position Conn To QDF27 22...

Page 186: ...212 MTA 2 Positions Standard Siderail Assembly Head End Left QDF27 1429 J1 To QDF27 1682 MTA 3 Positions QDF27 1429 J2 To QDF27 1206 MTA 3 Positions Standard Siderail Assembly Head End Left and Standard Siderail Assembly Head End Right QDF27 1682 MTA 3 Positions To QDF27 1429 J1 QDF27 1682 MTA 3 Positions To QDF27 1099 J4 Standard Siderail Assembly Head End Left and Standard Siderail Assembly Head...

Page 187: ... 2 VV23A9C12HL Pan Head Tapping Screw 20 3 VV83A9G16 Pan Head Tapping Screw 3 4 QDF27 2196 Head Speaker 2 5 QDF27 2198 Nurse Call and Speaker Fascia Right 1 6 QDF27 1429 Nurse Call Board 2 7 QDF27 2199 Nurse Call and Speaker Fascia Left 1 8 QDF27 2216 Siderail Speaker Right 1 9 QDF27 2175 Siderail Speaker Left 1 10 QDF27 1682 Siderail Nurse Call Wires 2 Optional Siderail Assembly Head End with Spe...

Page 188: ... and Standard Siderail Assembly Head End Right 27 2212 5 6 7 2 27 2212 QDF27 2211 QDF27 2210 OL270322 Rev B Reference Only NOTE 1 MAKE SURE THAT A JUMPER OR A CONNECTOR IS PLUGGED ON J14 TO J16 OF QDF75 0270 BOARD IF NO CONNECTOR IS PLUGGED PUT A JUMPER ON PINS 2 AND 3 FOR J14 TO J16 OF QDF75 0600 BOARD 3 9 SEE 27 2706 FOR DIP SWITCH CONFIGURATION AND NOTE 1 FOR CONNECTORS CONFIGURATION 12 10 13 N...

Page 189: ...OOM INTERFACE BOARD J5 TOWARDS ROOM INTERFACE BOARD J2 13 14 SEE NOTE 2 12 5 3 6 WHEN ZOOM OPTION IS INSTALLED 27 2706 TO SEE THE DIP SWITCH CONFIGURATION AND NOTE 1 FOR CONNECTORS CONFIGURATION SEE NOTE 2 16 15 5 OL270302 Rev B Reference Only QFD27 2199 QFD27 2198 Note 2 Attach to other cables with cable tie Item 14 Note 1 Make sure that a jumper or a connector is plugged on J14 to J16 of the QDF...

Page 190: ... iBed Wireless LEFT IR MODULE CABLE ITEM 13 SEE NOTE 3 4 CABLE FROM RIGHT IR MODULE ITEM 11 SEE NOTE 3 4 SEE NOTE 2 3 4 SEE NOTE 2 3 4 TOP VIEW HEAD FRAME COVER NOT SHOWN LEFT IR MODULE CABLE ITEM 13 CABLE FROM RIGHT IR MODULE ITEM 11 SEE NOTE 2 3 4 12 TOP VIEW HEAD FRAME COVER NOT SHOWN Path of IR Module Cables to the Room Interface Board ...

Page 191: ...m 2141 209 002 REV H 191 UNDER BED VIEW SEE NOTE 3 SEE NOTE 3 14 14 11 UNDER BED VIEW SEE NOTE 3 14 13 Optional Siderail Assembly Head End with Speaker with NC with iAudio with iBed Wireless Path of IR Module Cables to the Room Interface Board ...

Page 192: ...ight 1 14 QDF9518 Cable Tie 3 15 QP27 2735 IR Support Gasket Left 1 16 QP27 2734 IR Support Gasket Right 1 Cable No Connector No To Cable No Connector No QDF27 2175 MTA 2 Positions To QDF27 2212 MTA 2 Positions Standard Siderail Assembly Head End Left QDF27 1429 J1 To QDF27 1682 MTA 3 Positions QDF27 1429 J2 To QDF27 1206 MTA 3 Positions Standard Siderail Assembly Head End Left and Standard Sidera...

Page 193: ...41 209 002 REV H 193 2 3 4 1 Item Part No Part Name Qty 1 27 2289 Right Foot Siderail Assembly 1 2 27 2290 Left Foot Siderail Assembly 1 3 VB18A1O32 Bolt 8 4 VE30A1O Nylon Locknut 8 Siderail Mounting Assembly Foot End L27 032 Rev 02 Reference Only ...

Page 194: ...S FRAME 7 5 21 T35 3 5 15 24 23 12 20 19 4 23 14 18 3 13 17 6 26 16 9 1 22 A 6 36 SEE SIDERAIL OPTION 37 37 39 LUBRICATE B A 41 27 40 PART QP28 0039 IS OF GRAY COLOR CODE OF COLOR 10 T22 2 Standard Siderail Assembly Foot End Right 27 2289 Rev 03 Reference Only 1 DETAIL A 33 34 8 38 RACEWAY ...

Page 195: ...208 MTA 100 3 Positions QDF27 1521 MTA 100 3 Positions QDF27 1208 Mate N Lock 3 Positions QDF27 1481 J172 L27 029 POSITION ON CONNECTOR Cable QDF27 1208 7 Signal Color Position on Connector 37 A Position on Connector 37 B 1 NO SWITCH White 1 2 COM SWITCH Black 2 3 NC SWITCH Green 3 4 LED1 LBS Orange 1 5 12V Red 3 6 LED2 LBS Blue 2 Note 1 Siderail without iBed option See OL270059 XXX page 200 Sider...

Page 196: ...l Handle 1 14 QP27 1508 Right Foot Siderail Cap 1 15 QP28 0039 Siderail Rotation Arm Cover 2 16 QPA27 2297 Right Foot Siderail Structure 1 17 QPA28 0015 Siderail Sleeve 1 18 QPA28 0016 Sleeve with Lock 1 19 QRE27 1736 Return Spring 1 20 VV10A0G16 S Hexagon Cylindric Head Screw 2 21 VV10B0G40 S Hexagon Cylindric Head Screw 4 22 VV10A1P20 S Hexagon Cylindric Head Screw 1 23 VV23A9C12HL Phillips Pan ...

Page 197: ...Assembly Foot End Left 27 2290 Rev 03 Reference Only 1 1 1 B C B C 5 6 12 21 10 4 7 13 13 9 1 19 18 T35 3 5 15 17 20 TOWARDS FRAME D D 24 20 27 23 3 24 A 29 28 SEE SIDERAIL OPTION 29 A B 11 37 36 26 16 14 T22 2 PART QP28 0039 IS OF GRAY COLOR CODE OF COLOR 10 DETAIL A 8 33 34 35 RACEWAY ...

Page 198: ...08 MTA 100 3 Positions QDF27 1521 MTA 100 3 Positions QDF27 1208 Mate N Lock 3 Positions QDF27 1481 J171 L27 029 Note 1 Siderail without iBed option See OL270059 XXX page 200 Siderail with iBed option See OL270060 XXX page 201 POSITION ON CONNECTOR Cable QDF27 1208 24 Signal Color Position on Connector 29 A Position on Connector 29 B 1 NO SWITCH White 1 2 COM SWITCH Black 2 3 NC SWITCH Green 3 4 L...

Page 199: ...Left Siderail Casting 1 14 QPA28 0015 Siderail Sleeve 1 15 QPA28 0016 Sleeve with Lock 1 16 QRE27 1736 Return Spring 1 17 VV10A0G16 S Hexagon Cylindric Head Screw 2 18 VV10B0G40 S Hexagon Cylindric Head Screw 4 19 VV10A1P20 S Hexagon Cylindric Head Screw 1 20 VV23A9C12HL Phillips Pan Head Screw 15 21 VV31A0G16 Phillips Flat Head Screw 2 23 VVB3A9N24PF Bosscrew Screw 6 24 QDF27 1208 Foot Siderail L...

Page 200: ...Left Siderail Shown OL270059 XXX Rev 05 Reference Only NOTE XXX indicates language choice tri English French Spanish bil English French Item Part No Part Name Qty 1 QDF27 0909 XXX Right Foot Siderail Cover Fascia 1 2 QDF27 0910 XXX Left Foot Siderail Cover Fascia 1 3 QP27 1831 LBS Lens Foot Siderail 2 4 VV23A9C12HL Pan Head Tapping Screw 4 ...

Page 201: ...sh French Spanish bil English French Item Part No Part Name Qty 1 QDF27 2030 XXX Right Foot Siderail Fascia with iBed 1 2 QDF27 2031 XXX Left Foot Siderail Fascia with iBed 1 3 QDF27 1834 LBS Cable Foot Siderail 2 4 QDF2080 Inline Splice 2 5 QDF27 1562 iBED Electronic Board Lens 4 6 QP27 1831 LBS Foot Siderail Lens 2 7 VV23A9C12HL Pan Head Tapping Screw 4 Position on Wire 27 1834 Position on Conne...

Page 202: ...To Table of Contents 202 2141 209 002 REV H www stryker com Headboard and Footboard Assembly 1 L27 028 Rev E Reference Only Item Part No Part Name Qty 1 27 1547 Footboard Assembly 1 3 QP27 1077 Head Panel 1 3 ...

Page 203: ...dard Footboard Assembly 27 1547 Rev R Reference Only 2 3 1 11 9 19 8 7 17 6 15 14 20 20 A A 24 23 22 21 17 25 26 10 See Foot Control Options See Foot Control Options Note Pierce holes to install item 20 and make opening to root the cables if necessary See Note See Note 27 ...

Page 204: ...QP27 1493 Right Foot Control Support 1 9 QP27 1511 Foot Control Pivot Bushing 2 10 27 2785 Main Footboard 1 11 QP27 1569 Left Foot Control Support 1 14 VV23A1G24HL Phillips Screw 2 15 VV23A9C12HL Pan Head Tapping Screw 2 17 VV37A1N20 S Truss Head Machine Screw 4 19 QDF21 3943 Screw Cover 4 20 QDF9518 Cable Tie 2 21 VE30A1G Nylon Hex Locknut 2 22 VW10A06 Flat Washer 2 23 0056 028 000 Black TPR Bump...

Page 205: ...tents www stryker com 2141 209 002 REV H 205 Footboard Assembly without iBed 3 16 9 11 11 1 14 5 4 A 27 1547 6 7 8 19 T35 24 T5 23 20 10 24 T5 17 T5 17 T5 18 T15 15 19 T35 19 T35 27 26 28 21 OL270328 XXX Rev B Reference Only ...

Page 206: ...er com 2 13 17 T5 Footboard Assembly without iBed 22 12 BOARD ORIENTED TO HAVE THE WRITTING ON THE RIGHT SIDE VISIBLE 23 12 12 POSITION THE SCREEN TO HAVE THE YELLOW CABLES UPWARD 12 12 CONNECTION OF THE TOUCHSCREEN ON J3 BOARD QDF75 0290 BOARD ASSEMBLY 21 21 29 ...

Page 207: ...Spanish bil English French Footboard Assembly without iBed OL270328 XXX Rev B Reference Only Item Part No Part Name Qty 1 QDF27 1097 Brake Control Board 1 2 QDF27 2054 iBed without Lens Label 1 3 QDF27 2756 Back Foot Control Fascia 1 4 QDF27 2188 XXX Foot Control Overlay 1 5 QDF27 2193 8 in Touch Screen 1 6 QDF27 2229 Brake to Interface Cable 1 7 QDF27 2230 Menu to Interface Cable 1 8 QDF27 2232 F...

Page 208: ...ents 208 2141 209 002 REV H www stryker com 5 18 11 13 13 1 16 7 6 A 27 1547 28 3 8 9 10 21 T35 26 T5 25 22 12 21 T5 19 T5 19 T5 20 T15 17 21 T35 21 T35 29 30 23 Footboard Assembly with iBed OL270327 XXX Rev B Reference Only ...

Page 209: ...209 Footboard Assembly with iBed A 4 3 15 19 T5 2 22 25 14 14 14 14 CONNECTION OF THE TOUCHSCREEN ON J3 BOARD POSITION THE SCREEN TO HAVE THE YELLOW CABLES UPWARD 14 BOARD ORIENTED TO HAVE THE WRITTING ON THE RIGHT SIDE VISIBLE QDF75 0290 BOARD ASSEMBLY 23 31 31 23 ...

Page 210: ...NOTE XXX indicates language choice tri English French Spanish bil English French Footboard Assembly with iBed OL270327 XXX Rev B Reference Only Item Part No Part Name Qty 1 QDF27 1097 Brake Control Board 1 2 QDF27 1562 Local Bed Status Board 2 3 QDF27 2253 LBS Cable Foot Panel 1 4 QDF27 1610 iBed Lens Label 1 5 QDF27 2756 Back Foot Control Fascia 1 6 QDF27 2188 XXX Foot Control Overlay 1 7 QDF27 2...

Page 211: ...tryker com 2141 209 002 REV H 211 5 17 11 13 13 7 6 A 27 1547 3 8 9 10 16 19 T15 20 T35 26 25 18 T5 1 27 28 T5 29 30 19 2x T5 21 32 22 33 37 34 12 35 39 T5 36 23 Footboard Assembly with iBed and Wi Fi OL270326 XXX Rev B Reference Only ...

Page 212: ...board Assembly with iBed and Wi Fi A 3 2 15 4 18 T5 22 31 CONNECTION OF THE TOUCHSCREEN ON J3 BOARD 14 14 POSITION THE SCREEN TO HAVE THE YELLOW CABLES UPWARD 14 14 32 BOARD ORIENTED TO HAVE THE WRITTING ON THE RIGHT SIDE VISIBLE 14 36 QDF75 0290 BOARD ASSEMBLY 23 38 23 40 ...

Page 213: ...nglish French Spanish bil English French Footboard Assembly with iBed and Wi Fi OL270326 XXX Rev B Reference Only Item Part No Part Name Qty Item Part No Part Name Qty 1 QDF27 1097 Brake Control Board 1 2 QDF27 1562 Local Bed Status Board 2 3 QDF27 2253 LBS Cable Foot Panel 1 4 QDF27 1610 iBed Lens Label 1 5 QDF27 2756 Back Foot Control Fascia 1 6 QDF27 2188 XXX Foot Control Overlay 1 7 QDF27 2193...

Page 214: ...F64 1358 Bed Extender Cable 1 7 QE71 1350 XXX Manufacturer Sticker 1 8 64 1271P Bed Extender Protective Device 1 9 64 1278P Bed Extender Cover 1 10 QE71 1112 XXX Bed Extender Warning Label 1 11 64 1284P Bed Extender Connector Protective Device 1 12 QE71 0533 Protective Top Sticker 2 13 QDF6029 ABS Handle 2 14 QE71 1132 XXX Appropriate Cushion Label 1 15 VW10C081802 Nylon Washer 2 16 VE30A1E Nylon ...

Page 215: ...Return To Table of Contents www stryker com 2141 209 002 REV H 215 Bed Extender Pad with XPRT Mattress DM64197 XXX Bed Extender Pad with Position Pro Mattress DM64196 XXX 6 35 5 1 2 35 4 6 ...

Page 216: ...Return To Table of Contents 216 2141 209 002 REV H www stryker com Line Management Clip FA64210 XXX L64 130 XXX Rev A Reference Only ...

Page 217: ...w 1 11 VV81A9E16 10 Flat Head Tapping Screw 1 12 QE71 1350 XXX Manufacturer Sticker 1 13 VW10A10 Flat Washer 2 L64 139 XXX Rev A Reference Only 4 3 1 9 6 5 10 7 2 8 11 FA64221 XXX Left FA64238 XXX Right Item Part No Part Name Qty 1 QDF64 1366 I V Pole Support Right 1 2 QDF64 1367 Simple Stop I V Pole Left 1 3 QDF2121 Sleeve Nut 1 4 QDF5073 2 Stage I V Pole 1 5 QE71 1350 XXX Manufacturer Sticker 1 ...

Page 218: ...turn To Table of Contents 218 2141 209 002 REV H www stryker com IV Pole Ass y Dual Head End Permanent FA64202 XXX HEAD OF BED 8 9 4 11 5 5 6 13 15 14 12 TOQUE 75 10 3 2 1 L64 107 XXX Rev A Reference Only ...

Page 219: ...F64 1366 I V Pole Support Right 1 4 QP64 1225 I V Pole Plastic Support 2 5 QDF5073 2 Stage I V Pole 2 6 QE71 1306 XXX Warning Sticker 2 8 VB15A1O32 S Bolt 2 9 VD60A1N1016 S Shoulder Screw 2 11 VV81A9E16 10 Flat Head Tapping Screw 2 12 QE71 1350 XXX Manufacturer Sticker 1 13 VW10A10 Flat Washer 2 14 VV33A1G08 S Pan Head Machine Screw 4 15 64 1269C Double I V Pole Support 2 IV Pole Ass y Dual Head E...

Page 220: ...ts 220 2141 209 002 REV H www stryker com Monitor Tray Assembly FA64214 XXX 1 2 3 4 Item Part No Part Name Qty 1 QDF64 1300 Tray 1 2 QDF64 1301 Transport Strap 1 3 QDF64 1302 Fastener Sleeve 1 4 QDF64 1303T Tray Support Pole 1 ...

Page 221: ...4187 XXX Item Part No Part Name Qty 1 64 0647 Security Chain 1 2 QDF64 1059 Bottle Holder 1 3 QE71 1350 XXX Manufacturer Sticker 1 4 QE71 0601 Maximum Load Sticker 1 5 VR11H46 Dome Head Pop Rivet 1 6 64 1285 Oxygen Bottle Holder Socket 1 7 VG10B0628 Spring Pin 1 2 4 3 1 5 6 7 L64 112 XXX Rev A Reference Only ...

Page 222: ...Return To Table of Contents 222 2141 209 002 REV H www stryker com Right Fit Oxygen Bottle Holder Assembly FA64203 L64 124 Rev 00 Reference Only ...

Page 223: ...com 2141 209 002 REV H 223 Pendant Assembly with Motion Control FA64228 XXX L64 146 XXX Rev A Reference Only 2 1 Motion Control Item Part No Part Name Qty 1 QDF64 1240 Removable Pendant with Motion Control 1 2 QE14399 T Manufacturer Sticker 1 ...

Page 224: ...www stryker com Pendant Assembly with Motion Control NC FA64226 L64 118 Rev 2 Reference Only Item Part No Part Name Qty 1 QDF64 1238 Removable Pendant with Nurse Call and Motion Control 1 2 QE14399 T Manufacturer Sticker 1 2 1 Motion Control Nurse Call ...

Page 225: ... 002 REV H 225 Pendant Assembly with Motion Control Smart TV FA64227 L64 119 Rev 3 Reference Only 2 1 Motion Control Smart TV Item Part No Part Name Qty 1 QDF64 1239 Removable Pendant with Motion Control and Smart TV 1 2 QE14399 T Manufacturer Sticker 1 ...

Page 226: ...rt Name Qty 1 QE71 1350 XXX Manufacturer Sticker 1 2 QDF64 1282 Additional Removable Grip 1 Pendant Assembly with Motion Control NC Smart TV FA64225 L64 117 Rev 2 Reference Only 2 1 Motion Control Nurse Call Smart TV Item Part No Part Name Qty 1 QDF64 1237 Removable Pendant with Nurse Call Motion Control and Smart TV 1 2 QE14399 T Manufacturer Sticker 1 ...

Page 227: ...ers 2 1 8 7 6 3 5 Item Part No Part Name Qty 1 27 2046P Left Bumper Support 1 2 27 2056P Right Bumper Support 1 3 QPC27 2059 Wheel 1 3 4 Diameter 2 5 VE30A1N Nylon Locknut 2 6 VW10C122802 Nylon Washer 2 7 27 2235Z Bumper Spacer 2 8 VB15A1N44 Bolt 2 OL270050 Rev 04 Reference Only ...

Page 228: ... stryker com Traction Sleeve Assembly 4 in x 1 2 in FA64215 XXX 1 2 4 3 Item Part No Part Name Qty 1 64 1262C Traction Sleeve 4 x 1 2 4 2 QE71 1350 XXX Manufacturer Sticker 1 3 VB15A1N32 S Hex Bolt 4 4 VW10A08 Flat Washer 4 L64 134 XXX Rev A Reference Only ...

Page 229: ...02 REV H 229 Traction Sleeve Assembly 4 in x 3 4 in FA64216 XXX 1 3 2 4 Item Part No Part Name Qty 1 64 1263C Traction Sleeve 4 x 3 4 4 2 QE71 1350 XXX Manufacturer Sticker 1 3 VB15A1N32 S Hex Bolt 4 4 VW10A08 Flat Washer 4 L64 135 XXX Rev A Reference Only ...

Page 230: ... stryker com Traction Sleeve Assembly 8 in x 1 2 in FA64217 XXX 1 3 2 4 Item Part No Part Name Qty 1 64 1264C Traction Sleeve 8 x 1 2 4 2 QE71 1350 XXX Manufacturer Sticker 1 3 VB15A1N32 S Hex Bolt 4 4 VW10A08 Flat Washer 4 L64 136 XXX Rev A Reference Only ...

Page 231: ...02 REV H 231 Traction Sleeve Assembly 8 in x 3 4 in FA64218 XXX 1 3 2 4 Item Part No Part Name Qty 1 64 1265C Traction Sleeve 8 x 3 4 4 2 QE71 1350 XXX Manufacturer Sticker 1 3 VB15A1N32 S Hex Bolt 4 4 VW10A08 Flat Washer 4 L64 137 XXX Rev A Reference Only ...

Page 232: ...tryker com Traction Sleeve Assembly 6 1 2 in x 3 4 in FA64219 XXX 1 3 Item Part No Part Name Qty 1 64 1266C Traction Sleeve 6 1 2 x 3 4 4 2 QE71 1350 XXX Manufacturer Sticker 1 3 VB15A1N32 S Hex Bolt 4 4 VW10A08 Flat Washer 4 L64 138 XXX Rev A Reference Only ...

Page 233: ...ble of Contents www stryker com 2141 209 002 REV H 233 Wall Saver Cable Assembly FA64208 1 2 L64 129 Rev 03 Reference Only Item Part No Part Name Qty 1 QDF64 1371 Wall Saver 1 2 QE71 1313 T Sticker Manufacturing 1 ...

Page 234: ...urn To Table of Contents 234 2141 209 002 REV H www stryker com 3 2 11 9 13 8 4 11 1 6 12 10 14 7 5 X ray Cassette Holder Assembly FA64205 XXX L64 123 XXX Rev A Reference Only 4 81 0 65 5 81 12 48 1 90 2X ...

Page 235: ...60Z Cassette Support 1 5 64 1276 X Ray Cassette Support Spacer 2 6 QDF2093 X Ray Cassette Manual Tightening Screw 1 7 QE71 1350 XXX Manufacturer Sticker 1 8 QE71 1080 X Ray Cassette Handle Sticker 2 9 QP64 1185 X Ray Cassette for Plate 1 10 VE40A1N Cap Nut 1 11 VV83A9G16 Phillips Head Tapping Screw 8 12 VV83A9G24 Phillips Head Tapping Screw 2 13 VW10A06 Flat Washer 2 14 M0008 Threadlocker Medium S...

Page 236: ... H www stryker com Tube Support 1 3 4 2 3 2 1 4 Item Part No Part Name Qty 1 27 2518P Tube Support 2 2 QE71 0498 Triangular Sticker 4 3 QE71 1248 XXX Sticker to Drainage Tube 2 4 VV83A9G24 Pan Phillips Head Screw 4 OL270224 XXX Rev 01 Reference Only ...

Page 237: ...2141 209 002 REV H 237 Assembly Part number 27 2287 27 2288 Reference Only Item Recycling Material Code Important Information Qty A QDF27 1097 Brake Control Board 2 B QDF27 1099 Siderail Nurse Control Board 2 A B Recycling Passports 27 2287 shown ...

Page 238: ...238 2141 209 002 REV H www stryker com Assembly Part number OL270311 Reference Only Item Recycling Material Code Important Information Qty A QDF27 1429 Nurse Call Board 2 B QDF75 0270 Room Interface Board 1 A B Recycling Passports ...

Page 239: ...om 2141 209 002 REV H 239 Assembly Part number OL270302 Reference Only Item Recycling Material Code Important Information Qty A QDF75 0270 Room Interface Board 1 B 27 2661 Left IR Module 1 C 27 2662 Right IR Module 1 A B Recycling Passports C ...

Page 240: ...To Table of Contents 240 2141 209 002 REV H www stryker com Assembly Part number OL270032 Reference Only Item Recycling Material Code Important Information Qty A QDF2060 Smart TV Board 1 A Recycling Passports ...

Page 241: ...f Contents www stryker com 2141 209 002 REV H 241 Assembly Part number OL270263 XXX Reference Only Item Recycling Material Code Important Information Qty A QDF27 1099 Siderail Nurse Control Board 3 A Recycling Passports ...

Page 242: ...Table of Contents 242 2141 209 002 REV H www stryker com Assembly Part number 27 2547 XXX Reference Only A Recycling Passports Item Recycling Material Code Important Information Qty A 27 2548 Zoom Control Board 1 ...

Page 243: ...02 REV H 243 Recycling Passports Assembly Part number OL270327 XXX Item Recycling Material Code Important Information Qty A QDF75 0010 Main Menu Board 1 B QDF27 1097 Brake Control Board 1 C QDF75 0290 Touch Board Assembly 1 D 27 2480 LCD Display 1 A C B D ...

Page 244: ...rt number OL270326 XXX Reference Only Item Recycling Material Code Important Information Qty A QDF75 0010 Main Menu Board 1 B QDF75 0630 USB Wi Fi Board 1 C QDF27 1097 Brake Control Board 1 D QDF75 0290 Touch Board Assembly 1 E 27 2480 LCD Display 1 Recycling Passports B A C D E ...

Page 245: ... www stryker com 2141 209 002 REV H 245 Recycling Passports Assembly Part Number 27 2688 Reference Only Item Recycling Material Code Important Information Qty A QDF75 0440 DC Power Control Board 1 B QDF9188 12V 18Ah Battery 2 A B ...

Page 246: ...ts 246 2141 209 002 REV H www stryker com Recycling Passports Assembly Part number 27 2546 Reference Only Item Recycling Material Code Important Information Qty A QDF27 1430 Zoom Board 1 B QDF75 0240 Zoom Interface Board 1 A B ...

Page 247: ...n To Table of Contents www stryker com 2141 209 002 REV H 247 Recycling Passports Assembly Part number 27 2661 Reference Only Item Recycling Material Code Important Information Qty A QDF75 0310 IR Board 1 A ...

Page 248: ...n To Table of Contents 248 2141 209 002 REV H www stryker com Recycling Passports Assembly Part number 27 2662 Reference Only Item Recycling Material Code Important Information Qty A QDF75 0310 IR Board 1 A ...

Page 249: ...including any implied warranty of merchantability or fitness for a particular purpose are expressly excluded by Stryker In no event shall Stryker be liable for incidental or consequential damages TO OBTAIN PARTS AND SERVICE Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives These representatives are factory trained available locally and carry ...

Page 250: ......

Page 251: ......

Page 252: ...2006 XX XXXX XXX XXX REV X www stryker com 2018 03 2141 209 002 REV H www stryker com Manufactured For Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA ...

Reviews: