To
Signal Calibration
menu
To
Signal Calibration
menu
Calibration
- GPD/LPD Calibration
- Signal Calibration
- GO TO CONFIG MENU
Signal Calibration
- 4-20mA IN
- 0 - 10V IN
- 4-20mA OUT
To top of
Calibration
menu
Signal Calibration
- 0 - 10V IN
- 4-20mA OUT
- GO TO CAL MENU
Connect pump to 0-10V loop
using a process meter or powered transmitter
CONFIGURATION MENU
continued
CALIBRATION
page 2 of 2
Allows user to calibrate analog signal inputs and outputs.
0 - 10V IN Set
Input 0V
0 - 10V IN Set
Input 10V
set
back
set
Signal Calibration
- 4-20mA IN
- 0 - 10V IN
- 4-20mA OUT
When pump indicates to input
10V, bring loop to 10.0V (or
desired level). Press ENTER.
When pump indicates to input
0V, bring loop to 0.0V (or
desired level). Press ENTER.
Signal Calibration
- 4-20mA IN
- 0 - 10V IN
- 4-20mA OUT
Continued from previous page
Connect process meter to 4-20mA output signal
4mA OUT Set
0508
adjust
set
20mA OUT Set
2584
set
back
In 20mA Out Set, press the
UP
or
DOWN
button to adjust
the signal until the meter
reads 20.0. Press ENTER.
In 4mA Out Set, press the
UP
or
DOWN
button to adjust the
signal until the meter reads
4.0. Press ENTER.
Input 4mA
4-20mA IN Set
Input 20mA
set
back
set
Connect pump to 4-20mA loop
using a process meter or powered transmitter
4-20mA IN Set
When pump indicates to input
20mA, bring loop to 20.0mA
(or desired level). Press ENTER.
When pump indicates to input
4mA, bring loop to 4.0mA
(or desired level). Press ENTER.
To
Signal Calibration
menu
US and Canada 800.683.2378, International 904.641.1666
19
adjust
CAUTION:
4-20mA and
0-10VDC input signals
are pre-calibrated at
the factory. Only change
the input calibration if
the factory settings are
not sufficient.