36
Section 3
Medication Reservoir or Administration Set
Attaching a New Reservoir or Administration Set
Refer to the instructions for use supplied with the reservoir or
administration set for information on filling the reservoir, attaching a
CADD
®
extension set with anti-siphon valve or an add on anti-siphon valve,
and preparing the product for use.
The following sterile, single use products are compatible with the
CADD-Prizm
®
VIP pump:
>
Medication cassette reservoir (50 or 100 ml), used with the
CADD
®
extension set with anti-siphon valve
>
CADD
®
administration set with integral anti-siphon valve for
rates up to 125 ml/hr
>
CADD-Prizm
®
high volume administration set for rates up to
350 ml/hr (not for use in PCA delivery mode)
Rates above 125 ml/hr require a CADD-Prizm
®
high volume
administration set. Rates above 250 ml/hr also require an
AC adapter or a power pack.
>
CADD
®
tubing sets with flow stop free-flow protection
When you attach a reservoir or administration set, the pump’s screens
will lead you through the following steps:
>
Verifying the type of cassette
>
Resetting the reservoir volume (if the reservoir volume has
counted down)
>
Priming the fluid path (if the lock level is LL0 or LL1)
>
Changing the lock level (if AutoLock is not in use and
the lock level is LL0)
>
Starting the pump
CAUTION:
If you are using a CADD
™
medication cassette reservoir in which the
medication is frozen, thaw at room temperature only. Do not heat in a microwave oven
as this may damage the product and cause leakage.