25
Pump Programming - TPN Delivery Mode
Section 2
Pump
Pr
ogr
amming
TPN Programming
Before programming:
>
Stop the pump and change the Lock Level to LL0 as described on
page 9.
>
Select the TPN delivery mode as described in Changing Delivery
Modes on page 10. If the pump is already in the TPN delivery mode,
you may select it again to clear all programming screens to their
default settings.
1. Begin at the main screen
>
Make sure the pump is in LL0.
>
Make sure TPN and STOPPED appear
on the main screen.
>
Press
„
to begin.
2. Enter the reservoir volume
>
Press
´
or
Î
to select the volume
of fl uid in a fi lled reservoir. (If you do
not wish to use the reservoir volume
feature, scroll down to “Not In Use”
located before 10.0.)
>
Press
€
.
3. Enter the infusion volume
When this screen is fi rst displayed, the third line may show the
volume left from the last infusion period. As soon as you
begin scrolling to a new infusion volume, it will disappear.
>
Press
´
or
Î
to select the
desired volume of fl uid to be
delivered in milliliters.
>
Press
€
.
Reservoir Volume
| 1000 ml
<Range: 10.0 - 9990>
*** TPN 6230X ***
STOPPED
Press NEXT to advance
Infusion Volume
| 1000.0 ml total
<Range: 10 - 9990>
IN19822.indd 25
10/24/2011 3:57:14 PM