24
Section 2
Pump Programming - TPN Delivery Mode
TPN Cycle Graph and Programming Screens
The TPN delivery mode allows the high volume delivery of
solutions, with optional tapering. Delivery can be gradually
increased or “tapered up” at the beginning of the infusion profi le,
or it can be gradually decreased, or “tapered down” at the end of
the infusion profi le. An automatic Keep Vein Open (KVO) rate is
delivered at the end of the infusion profi le.
The following fi gure shows the programming screens in the
TPN delivery mode:
Reservoir Volume
Infusion Volume
Infusion Period
Taper-Up Period
Taper-Down Period
Rate
(Review)
Milliliters Given
Air Detector
(Review)
*** TPN 6230X ***
STOPPED
Press NEXT to advance
If Air Detector
is installed
Infusion Peroid
Infusion Volume
Taper-Up
Period
Taper-Down
Period
KVO
Infusion Profile
Plateau Rate
(must be at least 10 min.)
IN19822.indd 24
10/24/2011 3:57:14 PM
Summary of Contents for CADD-Prizm VIP 6100
Page 1: ...CADD Prizm VIP Ambulatory Infusion Pump Model 6100 and 6101 Clinician Information ...
Page 4: ...ii ...
Page 63: ......