37
Section 3
Medication Reservoir or Administration Set
Section 3
Reserv
oir or
Admin Set
To attach the new reservoir or administration set
Refer to the Instructions for Use with the reservoir
or set you will be attaching.
1. Clamp the tubing. Insert the cassette hooks
into the hinge pins on the pump.
2. Place the pump upright on a firm,
flat surface. Press down so the cassette
fits tightly against the pump.
3. Insert a coin or the side of the key into the latch
button, push in, and turn counterclockwise until
the mark on the latch lines up with the solid dot
and you feel the button click into place.
4. If appropriate, insert the pump key into the
lock and turn counterclockwise until the white
mark lines up with the solid dot.
Press
„
.
NOTE: In the PCA delivery mode, The cassette must be
locked in order to start the pump.
WARNING:
>
Attach the cassette properly. An improperly
attached or detached cassette could result in
unregulated gravity infusion of medication from
the fluid container or a reflux of blood.
>
You must use a CADD
®
extension set with
anti-siphon valve or a CADD
®
administration
set with either an integral or an add on
anti-siphon valve to protect against unregulated
gravity infusion that can result from an
improperly attached cassette.
D. Zurn 02/27/95
Prizm Latched & Locked
Lock
Latch
IN19822.indd 37
10/24/2011 3:57:18 PM