Section 2
Pump Programming - TPN Delivery Mode
4. Enter the infusion period
When this screen is fi rst displayed, the third line may show the
amount of time left from the last infusion period. As soon as you
begin scrolling to a new infusion period, it will disappear.
>
Press
´
or
Î
to select the period
over which infusion volume will be
delivered.
>
Press
€
.
5. Enter the taper-up period
>
Press
´
or
Î
to select the period
over which infusion will gradually
increase to plateau rate.
>
Press
€
.
6. Enter the taper-down period
>
Press
´
or
Î
to select the period
over which infusion will gradually
decrease to end of infusion period.
>
Press
€
.
7. View the calculated rate
>
Both the plateau rate and the KVO
rate will be displayed. (The KVO is
1/10 of the plateau rate or 5.0 ml/hr
maximum.) Press
„
to continue.
8. Clear the milliliters given
>
Press
€
if you wish to clear the
total amount delivered since the time
and date shown; even if the amount
is zero, this updates the time and
date markers.
Infusion Period
| 10 hrs 00 min
<Range: 3:00-99:50>
Taper-Up Period
| 1 hrs 30 min
<Range: 0:00- 9:50>
Taper-Down Period
| 1 hrs 30 min
<Range: 00:00- 8:20>
Milliliters Given
0.0 ml
since 01/08/96 10:35
Press ENTER to clear
Plateau Rate
117.6 ml/hr
KVO Rate 5.0 ml/hr
<Review Only>