35
Section 3
Medication Reservoir or Administration Set
Section 3
Reserv
oir or
Admin Set
Removing a Used Reservoir or Administration Set
WARNING:
Always close the fluid path tubing with the clamp before removing the
cassette from the pump to prevent unregulated gravity infusion.
1. Close the tubing clamp.
2. If the cassette is locked, insert the key and turn
the lock clockwise one-quarter turn until it stops.
(The mark on the lock will line up with open dot.)
3. Use a coin or the side of the key to unlatch
the cassette (the part of the reservoir or
administration set that attaches to the pump).
Insert the coin or the side of the key into the
slot and turn clockwise until the latching button
pops out. (The mark on the latch will line up
with open dot.)
4. Remove the cassette hooks from the
pump hinge pins.
2000-03-08 D. Zurn
ÇPrizm Remove CassÈ
IN19822.indd 35
10/24/2011 3:57:17 PM