27
Pump Programming - TPN Delivery Mode
Section 2
Pump
Pr
ogr
amming
9. Verify the air detector status
This screen will appear only if an air detector is installed.
>
Make sure the setting is correct.
NOTE: If the air detector is not required,
this screen will show whether it is turned
on or off.
>
Press
„
to continue. If you need to
correct the air detector setting, see
the Operator’s Manual.
10. Review the program
Press
„
repeatedly to review the programming screens.
If you need to reprogram a setting, press
„
until the
appropriate screen appears and change the setting as described
in this section.
Air Detector
Required
<Review Only>