54
Section 6
Reference & Troubleshooting
Install a fresh 9-volt battery. The pump will
not start with a depleted 9-volt battery.
A good battery must always be installed
even when an external source of power is
connected.
NOTE: This message may appear when you install
a fresh battery while an external source of power
is connected. Remove and reinstall the battery
to cancel this message, then restart the pump if
necessary.
The 9-volt battery is low but the pump is
operable. Change the 9-volt battery soon.
NOTE: This message may appear when you install
a fresh battery while an external source of power
is connected. Remove and reinstall the battery to
cancel this message.
The air detector is being turned on
automatically because the pump is
powering up in Lock Level 0. Use Options
if you wish to turn off the air detector.
The air detector has detected air in the fluid
path; the fluid path may contain air bubbles,
or the tubing may not be threaded through
the air detector. Press
„
to silence the
alarm, then:
> Make sure the tubing is threaded
properly.
> If the fluid path contains air bubbles,
close the clamps and disconnect the
fluid path from the patient. Then follow
the instructions for removing air using
the Prime Option on page 43.
Alarms and Messages, Alphabetical List
A more detailed list of alarms and messages is included in the
Operator’s Manual.
Message
Corrective Action
-
9-volt Battery Depleted /
Install Good Battery
9-volt Battery Low
Air Detector is Being
Turned On
Air In Line Detected /
Pump Will Not Run