Section 6
Reference & Troubleshooting
The power pack is depleted and unable to
support pump operation. The 9-volt battery
is supplying power. Recharge the power
pack with the AC adapter.
Your institution may have established a
maintenance program for the pump and
preventive maintenance is due. Refer to your
institution’s policy.
This message appears on rate, dose, or
reservoir volume screens when the pump
is in LL1. It indicates that the range of
programmable values is limited by the
value programmed in LL0 (i.e. you cannot
increase the value beyond what was
programmed in LL0).
The reservoir volume has reached 0.0 ml.
Press
„
to stop the alarm. Then install a
new reservoir if appropriate.
The reservoir volume value is low,
indicating that the level of fluid in the reservoir
is low. Prepare to install a new reservoir.
Fluid is not flowing from the reservoir to
the pump. Check for a kink, a closed clamp
or an air bubble in the tubing between
reservoir and pump. Press
⁄
to stop
the pump and silence the alarm for two
minutes, then remove the obstruction and
press
„
to restart the pump.
The pump detects the cassette is damaged,
attached improperly, or incompatible with the
pump. Close the tubing clamp. Make sure the
cassette is attached properly. Then open the
clamp and restart the pump. If the alarm
persists, you may need to replace the cassette.
Message
Corrective Action
Power Pack Depleted /
Change Power Source
Prev. Maint. Reminder
(Date)
Range: Limited
Reservoir Volume is Zero
Reservoir Volume Low
Upstream Occlusion
(Model 6101 Only)
Wrong Cassette