31
Pump Programming - Intermittent Delivery Mode
Section 2
Pump
Pr
ogr
amming
5. Enter the dose cycle
>
Press
´
or
Î
to select the time
from the start of one dose to the start
time of the next dose.
>
Press
€
.
6. Enter the KVO rate
>
Press
´
or
Î
to select the amount
of drug delivered to maintain catheter
patency.
>
Press
€
.
7. Enter the next dose start time
>
Press
´
or
Î
to select the date and
time designated to start delivery of
next dose (or “Immediate”).
>
Press
€
.
>
Verify the start time shown on the
screen and press
´
if correct.
If you need to re-enter the start time
and date, press
Î
.
8. Verify the dose rate
>
The pump automatically calculates
the rate of dose delivery based on
the volume and duration you entered.
This screen is for review only; press
„
to continue. (If the rate is
125 ml/hr or greater, a CADD-Prizm
®
high volume administration set must
be used.)
9. Clear the milliliters given
>
Press
€
if you wish to clear the
amount given; even if the amount is
zero, this updates the time and date
markers.
Dose Cycle
| 6 hrs 00 min
<Range: 01:05-96:00>
KVO Rate
| 0.2 ml/hr
<Range: 0.0 - 10>
Next Dose Start Time
| 20:00 2/21/96
<Range: See Help>
Start next dose in
3 hrs 27 min on
2/21/96 at 20:00?
Press Y or N
Milliliters Given
0.0 ml
since 01/08/96 10:35
Press ENTER to clear
Dose Rate
50.0 ml/hr
<Review Only>